A DISPOSABLE END-TIDAL CO2 DETECTOR TO VERIFY ENDOTRACHEAL INTUBATION

被引:55
|
作者
ANTON, WR
GORDON, RW
JORDAN, TM
POSNER, KL
CHENEY, FW
机构
[1] Respiratory Care Services, University of Washington Medical Center, Seattle
[2] Department of Anesthesiology, University of Washington Medical Center, Seattle
关键词
end-tidal carbon dioxide; endotracheal intubation;
D O I
10.1016/S0196-0644(05)80938-2
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Study objective: We compared the performed of the Fenem FEF(TM) endtidal CO2 detector with TRIMED capnometer to verify endotracheal intubation. Design: The FEF(TM) indicates the presence of CO2 by the color change of a chemically treated indicator; the TRIMED uses infrared technology. Both devices were used during 60 intubations. Setting: Intubations during in-hospital emergency situations outside of the operating room were studied. Type of participants: Adult patients undergoing intubation for respiratory failure, CPR, and other airway protection situations were enrolled in the study. Interventions: The TRIMED monitor and FEF(TM) detector were placed in series between the manual resuscitator and the patient's endotracheal tube adapter after endotracheal tube placement. Measurements and main results: We defined the acceptable criterion for detection of CO2 as production of a positive signal within six mannual resuscitator bag breaths. The TRIMED met this criterion in 58 of 60 patients (sensitivity, 0.97) and the FEF(TM) met this criterion in 59 of 60 patients (sensitivity, 0.98). A paired t test showed no statistically significant difference in performance between the two devices. In five of nine cases of intubation during CPR, the color change of the FEF(TM) was described as "subtle." In one CPR case, a positive signal was not obtained by either device. Conclusion: We conclude that the performance of the FEF(TM) CO2 detector is equal to that of the TRIMED monitor for verification of endotracheal intubation in nonCPR situations. Interpretation of FEF(TM) color changes during CPR should be approached with caution until further studies using the FEF(TM) during CPR are completed.
引用
收藏
页码:271 / 275
页数:5
相关论文
共 41 条
  • [21] Confirming gastric tube placement in neurocritical patients using the combined effects of end-tidal CO2 and pH testing: A prospective diagnostic study
    Chen, Yunxia
    Xu, Mengmin
    Cheng, Yanwei
    Li, Shuaishuai
    Gao, He
    Kong, Sha
    Zhang, Chi
    Hao, Xiaodi
    Yang, Mengli
    INTENSIVE AND CRITICAL CARE NURSING, 2025, 88
  • [22] A role of end-tidal CO2 monitoring for assessment of tracheal intubations in very low birth weight infants during neonatal resuscitation at birth
    Hosono, Shigeharu
    Inami, Ikuhiro
    Fujita, Hidetoshi
    Minato, Michiyoshi
    Takahashi, Shigeru
    Mugishima, Hideo
    JOURNAL OF PERINATAL MEDICINE, 2009, 37 (01) : 79 - 84
  • [23] UTILITY OF AN END-TIDAL CARBON-DIOXIDE DETECTOR DURING STABILIZATION AND TRANSPORT OF CRITICALLY ILL CHILDREN
    BHENDE, MS
    THOMPSON, AE
    ORR, RA
    PEDIATRICS, 1992, 89 (06) : 1042 - 1044
  • [24] Focus on POCUS: Identification of Early Successful Intubation by Point-of-Care Ultrasound Versus End-Tidal Carbon Dioxide: A Prospective Comparative Study
    Neethirajan, Soma Ganesh Raja
    Baskar, Ganeshamoorthy
    Parameswari, Aruna
    TURKISH JOURNAL OF ANAESTHESIOLOGY AND REANIMATION, 2024, 52 (06) : 240 - 246
  • [26] Assessing the cost-effectiveness of capnography for end-tidal CO2 monitoring during in-hospital cardiac arrest: A middle-income country perspective analysis
    Decker, Sergio Renato da Rosa
    Marzzani, Lucas Emanuel
    Ferreira, Pedro Rotta de
    Rosa, Paulo Ricardo Mottin
    Brauner, Janete Salles
    Rosa, Regis Goulart
    Bertoldi, Eduardo Gehling
    AMERICAN HEART JOURNAL PLUS: CARDIOLOGY RESEARCH AND PRACTICE, 2024, 40
  • [27] End-tidal carbon dioxide monitoring in emergency medicine, part 2: Clinical applications
    Ward, KR
    Yealy, DM
    ACADEMIC EMERGENCY MEDICINE, 1998, 5 (06) : 637 - 646
  • [28] End tidal CO2 and cerebral oximetry for the prediction of return of spontaneous circulation during cardiopulmonary resuscitation
    Engel, Thomas W., II
    Thomas, Craig
    Medado, Patrick
    Bastani, Aveh
    Reed, Brian
    Millis, Scott
    O'Neil, Brian J.
    RESUSCITATION, 2019, 139 : 174 - 181
  • [29] End tidal CO2 is reduced during hypotension and cardiac arrest in a rat model of massive pulmonary embolism
    Courtney, DM
    Watts, JA
    Kline, JA
    RESUSCITATION, 2002, 53 (01) : 83 - 91
  • [30] COMPARISON OF THE END-TIDAL ARTERIAL PCO(2) GRADIENT DURING EXERCISE IN NORMAL SUBJECTS AND IN PATIENTS WITH SEVERE COPD
    LIU, Z
    VARGAS, F
    STANSBURY, D
    SASSE, SA
    LIGHT, RW
    CHEST, 1995, 107 (05) : 1218 - 1224