TOTAL KNEE REPLACEMENT - RADIOLOGIC FINDINGS IN FAILURE OF POROUS-COATED METAL-BACKED PATELLAR COMPONENT

被引:8
作者
PIRAINO, D [1 ]
RICHMOND, B [1 ]
FREED, H [1 ]
BELHOBEK, G [1 ]
SCHILS, J [1 ]
STULBERG, B [1 ]
机构
[1] CLEVELAND CLIN FDN, DEPT ORTHOPAED SURG, CLEVELAND, OH 44195 USA
关键词
D O I
10.2214/ajr.155.3.2117356
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
One technique for replacing the articular surface of the patella in total knee replacement is the use of a cementless porous-coated metal-backed polyethylene component. Anteroposterior, lateral, and Merchant or skyline radiographs in 10 cases of porous-coated metal-backed components that failed were evaluated for component alignment, component displacement, radiolucencies, loose beads, and integrity of the patellar component. Component failure was documented by surgery in all 10 cases. Seven cases of failure involved the patellar pegs or junction of the pegs with the metal backing and three cases involved the polyethylene portion. In the seven cases of peg-metal backing failure, three cases showed loose beads before failure and seven cases showed displacement of the metal backing at time of failure. All three cases of polyethylene failure showed narrowing or displacement of the polyethylene portion. Our experience shows that displacement of the metal backing, displacement of the polyethylene, and disruption of patellar pegs are seen in failed porous-coated metal-backed patellar components. Loose beads may be seen before peg-metal backing failure.
引用
收藏
页码:555 / 558
页数:4
相关论文
共 13 条
  • [1] FAILURE OF THE METAL-BACKED PATELLAR COMPONENT AFTER TOTAL KNEE REPLACEMENT
    BAYLEY, JC
    SCOTT, RD
    EWALD, FC
    HOLMES, GB
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (05) : 668 - 674
  • [2] BAYLEY JC, 1988, CLIN ORTHOP RELAT R, P82
  • [3] BRICK GW, 1988, CLIN ORTHOP RELAT R, P163
  • [4] LOOSENING OF THE POROUS COATING IN TOTAL KNEE REPLACEMENT
    CHENG, CL
    GROSS, AE
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1988, 70 (03): : 377 - 381
  • [5] GOMES LSM, 1988, CLIN ORTHOP RELAT R, P72
  • [6] HUNGERFORD DS, 1983, CLIN ORTHOP RELAT R, P95
  • [7] LANDON GC, 1986, CLIN ORTHOP RELAT R, P49
  • [8] FRACTURE DISSOCIATION OF THE POLYETHYLENE IN METAL-BACKED PATELLAR COMPONENTS IN TOTAL KNEE ARTHROPLASTY
    LOMBARDI, AV
    ENGH, GA
    VOLZ, RG
    ALBRIGO, JL
    BRAINARD, BJ
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (05) : 675 - 679
  • [9] ROENTGENOGRAPHIC ANALYSIS OF PATELLOFEMORAL CONGRUENCE
    MERCHANT, AC
    MERCER, RL
    JACOBSEN, RH
    COOL, CR
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1974, A 56 (07) : 1391 - 1396
  • [10] Ranawat C.S., 1986, CLIN ORTHOP RELAT R, P93, DOI 10.1097/00003086-198604000-00012