Comparison of 21-gauge and 22-gauge Needles for Endobronchial Ultrasound-Guided Transbronchial Needle Aspiration of Mediastinal and Hilar Lymph Nodes

被引:39
作者
Saji, Junko [1 ]
Kurimoto, Noriaki [2 ]
Morita, Katsuhiko [2 ]
Nakamura, Miho [1 ]
Inoue, Takeo [1 ]
Nakamura, Haruhiko [2 ]
Miyazawa, Teruomi [1 ]
机构
[1] St Marianna Univ, Sch Med, Div Resp & Infect Dis, Dept Internal Med, Kawasaki, Kanagawa, Japan
[2] St Marianna Univ, Sch Med, Div Chest Surg, Dept Surg, Kawasaki, Kanagawa, Japan
关键词
EBUS-TBNA; needle; diagnostic yield; lung cancer;
D O I
10.1097/LBR.0b013e3182273b41
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
Background: Although endobronchial ultrasound-guided transbronchial needle aspiration (EBUS-TBNA) using a 22-gauge needle has emerged as an accurate, minimally invasive, and safe technique for accessing undiagnosed mediastinal adenopathy, particularly in patients with lung cancer, the small sample size obtained by a conventional needle may limit the accuracy of the result. Objectives: We evaluated the safety and efficacy of obtaining specimens using 21-gauge needles, comparing diagnostic yield with conventional 22-gauge needles. Methods: Participants consisted of 56 consecutive patients with mediastinal lesions referred to our institution for diagnostic EBUS-TBNA. The 21-gauge needle group (21 G) included 24 patients with 6 nonmalignancies and the 22-gauge needle group (22G) included 32 patients with 9 nonmalignancies. Final diagnosis was based on cytology, histology, surgical results, appropriate clinical pictures on examination, and/or clinical follow-up. Results: Comparing 21G and 22G, inadequate material rates were 0% versus 3.1% in cytology, and 4.2% versus 18.8% in histology, respectively. Accuracy in cytology, histology, and combined cytology and/or histology were 91.7% versus 65.6% (P=0.02), 95.8% versus 81.3% (P=0.11), and 100% versus 84.4% (P=0.04), respectively. After limiting cases to suspected malignancies, sensitivity in cytology, histology, and combined cytology and/or histology were 88.9% versus 52.2% (P=0.01), 100% versus 82.6% (P=0.09), and 100% versus 87.0% (P=0.17), respectively. Conclusions: Increasing sample volume using a 21-gauge needle rather than a 22-gauge needle might improve diagnostic yield in EBUS-TBNA. This study revealed the benefits of using a 21-gauge needle for cytological and histologic diagnostic yields.
引用
收藏
页码:239 / 246
页数:8
相关论文
共 18 条
[11]   Endobronchial and endoscopic ultrasound-guided real-time fine-needle aspiration for mediastinal staging [J].
Rintoul, RC ;
Skwarski, KM ;
Murchison, JT ;
Wallace, WA ;
Walker, WS ;
Penman, ID .
EUROPEAN RESPIRATORY JOURNAL, 2005, 25 (03) :416-421
[12]   COMPARISON OF THE WANG 19-GAUGE AND 22-GAUGE NEEDLES IN THE MEDIASTINAL STAGING OF LUNG-CANCER [J].
SCHENK, DA ;
CHAMBERS, SL ;
DERDAK, S ;
KOMADINA, KH ;
PICKARD, JS ;
STROLLO, PJ ;
LEWIS, RE ;
PATEFIELD, AJ ;
HENDERSON, JH ;
TOMSKI, SM ;
MORALES, CF ;
STERLING, JL ;
SOLANKI, PH ;
MOORE, J .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1993, 147 (05) :1251-1258
[13]   Yield of transbronchial needle aspiration in diagnosis of mediastinal lesions [J].
Sharafkhaneh, A ;
Baaklini, W ;
Gorin, AB ;
Green, L .
CHEST, 2003, 124 (06) :2131-2135
[14]   A Randomized Controlled Trial of Standard vs Endobronchial Ultrasonography-Guided Transbronchial Needle Aspiration in Patients With Suspected Sarcoidosis [J].
Tremblay, Alain ;
Stather, David R. ;
MacEachern, Paul ;
Khalil, Moosa ;
Field, Stephen K. .
CHEST, 2009, 136 (02) :340-346
[15]   FLEXIBLE TRANS-BRONCHIAL NEEDLE ASPIRATION FOR STAGING OF BRONCHOGENIC-CARCINOMA [J].
WANG, KP ;
BROWER, R ;
HAPONIK, EF ;
SIEGELMAN, S .
CHEST, 1983, 84 (05) :571-576
[16]   TRANS-BRONCHIAL NEEDLE ASPIRATION FOR DIAGNOSIS OF LUNG-CANCER [J].
WANG, KP ;
MARSH, BR ;
SUMMER, WR ;
TERRY, PB ;
EROZAN, YS ;
BAKER, RR .
CHEST, 1981, 80 (01) :48-50
[17]   Endobronchial ultrasound: new insight for the diagnosis of sarcoidosis [J].
Wong, M. ;
Yasufuku, K. ;
Nakajima, T. ;
Herth, F. J. F. ;
Sekine, Y. ;
Shibuya, K. ;
Iizasa, T. ;
Hiroshima, K. ;
Lam, W. K. ;
Fujisawa, T. .
EUROPEAN RESPIRATORY JOURNAL, 2007, 29 (06) :1182-1186
[18]   Real-time endobronchial ultrasound-guided transbronchial needle aspiration of mediastinal and hilar lymph nodes [J].
Yasufuku, K ;
Chiyo, M ;
Sekine, Y ;
Chhajed, PN ;
Shibuya, K ;
Iizasa, T ;
Fujisawa, T .
CHEST, 2004, 126 (01) :122-128