Comparison of 21-gauge and 22-gauge Needles for Endobronchial Ultrasound-Guided Transbronchial Needle Aspiration of Mediastinal and Hilar Lymph Nodes

被引:39
作者
Saji, Junko [1 ]
Kurimoto, Noriaki [2 ]
Morita, Katsuhiko [2 ]
Nakamura, Miho [1 ]
Inoue, Takeo [1 ]
Nakamura, Haruhiko [2 ]
Miyazawa, Teruomi [1 ]
机构
[1] St Marianna Univ, Sch Med, Div Resp & Infect Dis, Dept Internal Med, Kawasaki, Kanagawa, Japan
[2] St Marianna Univ, Sch Med, Div Chest Surg, Dept Surg, Kawasaki, Kanagawa, Japan
关键词
EBUS-TBNA; needle; diagnostic yield; lung cancer;
D O I
10.1097/LBR.0b013e3182273b41
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
Background: Although endobronchial ultrasound-guided transbronchial needle aspiration (EBUS-TBNA) using a 22-gauge needle has emerged as an accurate, minimally invasive, and safe technique for accessing undiagnosed mediastinal adenopathy, particularly in patients with lung cancer, the small sample size obtained by a conventional needle may limit the accuracy of the result. Objectives: We evaluated the safety and efficacy of obtaining specimens using 21-gauge needles, comparing diagnostic yield with conventional 22-gauge needles. Methods: Participants consisted of 56 consecutive patients with mediastinal lesions referred to our institution for diagnostic EBUS-TBNA. The 21-gauge needle group (21 G) included 24 patients with 6 nonmalignancies and the 22-gauge needle group (22G) included 32 patients with 9 nonmalignancies. Final diagnosis was based on cytology, histology, surgical results, appropriate clinical pictures on examination, and/or clinical follow-up. Results: Comparing 21G and 22G, inadequate material rates were 0% versus 3.1% in cytology, and 4.2% versus 18.8% in histology, respectively. Accuracy in cytology, histology, and combined cytology and/or histology were 91.7% versus 65.6% (P=0.02), 95.8% versus 81.3% (P=0.11), and 100% versus 84.4% (P=0.04), respectively. After limiting cases to suspected malignancies, sensitivity in cytology, histology, and combined cytology and/or histology were 88.9% versus 52.2% (P=0.01), 100% versus 82.6% (P=0.09), and 100% versus 87.0% (P=0.17), respectively. Conclusions: Increasing sample volume using a 21-gauge needle rather than a 22-gauge needle might improve diagnostic yield in EBUS-TBNA. This study revealed the benefits of using a 21-gauge needle for cytological and histologic diagnostic yields.
引用
收藏
页码:239 / 246
页数:8
相关论文
共 18 条
[1]   RAPID ON-SITE EVALUATION OF TRANSBRONCHIAL ASPIRATES [J].
DAVENPORT, RD .
CHEST, 1990, 98 (01) :59-61
[2]   The utility of transbronchial needle aspiration in the staging of bronchogenic carcinoma [J].
Harrow, EM ;
Abi-Saleh, W ;
Blum, J ;
Harkin, T ;
Gasparini, S ;
Addrizzo-Harris, DJ ;
Arroliga, AC ;
Wight, G ;
Mehta, AC .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2000, 161 (02) :601-607
[3]   FINE NEEDLE ASPIRATION CYTOLOGY - AN INVITRO STUDY OF CELL YIELD [J].
HARTLEY, MN ;
TUFFNELL, DJ ;
HUTTON, JL ;
PALMER, M ;
ALJAFARI, MS .
BRITISH JOURNAL OF SURGERY, 1988, 75 (04) :380-381
[4]   Real-time endobronchial ultrasound guided transbronchial needle aspiration for sampling mediastinal lymph nodes [J].
Herth, F. J. F. ;
Eberhardt, R. ;
Vilmann, P. ;
Krasnik, M. ;
Ernst, A. .
THORAX, 2006, 61 (09) :795-798
[5]   Actual role of endobronchial ultrasound (EBUS) [J].
Herth, Felix J. F. ;
Eberhardt, R. .
EUROPEAN RADIOLOGY, 2007, 17 (07) :1806-1812
[6]   Accuracy of transbronchial needle aspiration for mediastinal staging of non-small cell lung cancer: a meta-analysis [J].
Holty, JEC ;
Kuschner, WG ;
Gould, MK .
THORAX, 2005, 60 (11) :949-955
[7]   ENDOBRONCHIAL SONOGRAPHY - FEASIBILITY AND PRELIMINARY-RESULTS [J].
HURTER, T ;
HANRATH, P .
THORAX, 1992, 47 (07) :565-567
[8]   SAMPLE-SIZE IN FINE NEEDLE ASPIRATION BIOPSY [J].
KREULA, J ;
BONDESTAM, S ;
VIRKKUNEN, P .
BRITISH JOURNAL OF SURGERY, 1989, 76 (12) :1270-1272
[9]   Targeting Area in Metastatic Lymph Nodes in Lung Cancer for Endobronchial Ultrasonography-guided Transbronchial Needle Aspiration [J].
Kurimoto, Noriaki ;
Osada, Hiroaki ;
Miyazawa, Teruomi ;
Nishisaka, Takashi ;
Murayama, Masaki .
JOURNAL OF BRONCHOLOGY & INTERVENTIONAL PULMONOLOGY, 2008, 15 (03) :134-138
[10]   Comparison of 21-gauge and 22-gauge aspiration needle during endobronchial ultrasound-guided transbronchial needle aspiration [J].
Nakajima, Takahiro ;
Yasufuku, Kazuhiro ;
Takahashi, Ryo ;
Shingyoji, Masato ;
Hirata, Tetsushi ;
Itami, Makiko ;
Matsui, Yukiko ;
Itakura, Meiji ;
Iizasa, Toshihiko ;
Kimura, Hideki .
RESPIROLOGY, 2011, 16 (01) :90-94