THE COMPARISON OF INTRAOCULAR PRESSURE MEASUREMENT BY GOLDMANN APPLANATION TONOMETER AND NON-CONTACT AIR PUFF TONOMETER: A CROSS-SECTIONAL RESEARCH

被引:0
|
作者
Arooj, Bushra [1 ]
Nadeem, Asmarah [1 ]
Ali, Sobia [1 ]
机构
[1] Lahore Gen Hosp, Lahore, Pakistan
来源
INDO AMERICAN JOURNAL OF PHARMACEUTICAL SCIENCES | 2018年 / 5卷 / 12期
关键词
Tonometry; Glaucoma; Goldmann Applanation Tonometer and Air Puff Tonometer;
D O I
10.5281/zenodo.2526307
中图分类号
R914 [药物化学];
学科分类号
100701 ;
摘要
Objective: The aim of this particular research was to determine the accuracy and frequency of intraocular pressure which measured with the help of an Air Puff (non-contact tonometer); moreover, we also confirmed the same by Goldmann applanation tonometer. Material and methods: Our research design was cross-sectional and it was carried out at Ophthalmology Department of Mayo Hospital, Lahore (February to July 2017) on a total of 240 patients. All the patients were in the age bracket of (10 - 70) years including males and females. The selection process included those patients who visited Eye OPD for regular intraocular pressure, refraction, glaucoma, cataract surgery and suspected glaucoma. A complete ocular assessment was mandatory for every individual participating in the research. Results: In the age bracket of (10 - 70) years the mean age was (43.844 +/- 15.434) years and the values of mean IOP was for Goldmann Applanation Tonometer as (18.92 +/- 8.852) mmHg and for Air Puff Tonometer as (21.463 +/- 9.456) mmHg. The different measurement ranges for Air-Puff Tonometer and Goldmann Applanation Tonometer was respectively (10 - 54) mmHg and (11 - 58) mmHg. Conclusion: In order to measure intraocular pressure, the method of Air puff tonometer is non-contact and quick. Air Puff tonometer is effective for screening purposes; whereas, the intraocular pressure measurements are accurately measured through Goldmann applanation tonometer.
引用
收藏
页码:16875 / 16879
页数:5
相关论文
共 50 条
  • [31] Intraocular pressure difference in goldmann applanation tonometry versus a transpalpebral tonometer TGDc-01 "PRA" in glaucoma patients
    Schlote, T
    Landenberger, H
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2005, 222 (02) : 123 - 131
  • [32] A comparison of intraocular pressure measurement using SUOER SW-500 rebound tonometer and conventional reusable Goldmann prisms
    Chaung, Jia Quan
    Sangapillai, Thanendthire
    Quilat, Karen Kate
    Perera, Shamira
    FRONTIERS IN MEDICINE, 2024, 11
  • [33] Comparison of intraocular pressure measurements using Goldmann tonometer, I-care pro, Tonopen XL, and Schiotz tonometer in patients after Descemet stripping endothelial keratoplasty
    Ohana, Oded
    Varssano, David
    Shemesh, Gabi
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2017, 65 (07) : 579 - 583
  • [34] Evaluation of the applanation tonometer Tono-Pen Avia® Vet™ for the measurement of the canine and feline intraocular pressure
    Helbig, Stefanie
    Eule, J. Corinna
    TIERAERZTLICHE PRAXIS AUSGABE KLEINTIERE HEIMTIERE, 2021, 49 (02): : 86 - 92
  • [35] Comparison of intraocular pressure measurement with Scheimpflug-based noncontact tonometer with and without hydrogel contact lenses
    Kumar, Mukesh
    Shetty, Rohit
    Jayadev, Chaitra
    Dutta, Debarun
    Nicolsan, Maneck D.
    Nagaraj, Sriharsha
    Kumar, Rajesh S.
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2015, 63 (04) : 323 - 326
  • [36] Comparisons of the Tono-Pen® and Goldmann Applanation Tonometer in the Measurement of Intraocular Pressure of Primary Open Angle Glaucoma Patients in a Hospital Population in Southwest Nigeria
    Onochie, Chizoba
    Okoye, Obiekwe
    Ogunro, Adunola
    Aribaba, Tade
    Hassan, Kunle
    Onakoya, Adeola
    MEDICAL PRINCIPLES AND PRACTICE, 2016, 25 (06) : 566 - 571
  • [37] Comparability analysis of the HNT-1P Huvitz non contact tonometer for the measurement of intraocular pressure
    Montoya, Alicia Carceles
    Lucas, Carlos Enrique Monera
    Jaimes, Claudia Patricia Tarazona
    Valero, Daniel Romero
    Martinez, Alejandro Moya
    Toldos, Jose Juan Martinez
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2025, 35 (02) : 544 - 550
  • [38] New Non-contact Tonometer HNT-1P Reliability: Comparing Intraocular Pressure, Central Corneal Thickness, and Corrected Intraocular Pressure
    Eah, Kyu Sang
    Shin, Joong Won
    Sung, Kyung Rim
    JOURNAL OF THE KOREAN OPHTHALMOLOGICAL SOCIETY, 2020, 61 (05): : 524 - 531
  • [39] Comparison of intraocular pressure adjusted by central corneal thickness or corneal biomechanical properties as measured in glaucomatous eyes using noncontact tonometers and the Goldmann applanation tonometer
    Yaoeda, Kiyoshi
    Fukushima, Atsushi
    Shirakashi, Motohiro
    Fukuchi, Takeo
    CLINICAL OPHTHALMOLOGY, 2016, 10 : 829 - 834
  • [40] Intraocular pressure and corneal thickness - A comparison between non-contact tonometry and applanation tonometry
    Domke, N.
    Hager, A.
    Wiegand, W.
    OPHTHALMOLOGE, 2006, 103 (07): : 583 - 587