MEASUREMENT OF SERUM LP(A) BY COBAS MIRA USING A LATEX IMMUNOTURBIDIMETRIC ASSAY KIT

被引:7
|
作者
FUJITA, S [1 ]
SANO, T [1 ]
KATAYAMA, Y [1 ]
机构
[1] NATL CARDIOVASC CTR,DEPT CLIN CHEM,OSAKA,JAPAN
关键词
LIPID; LIPOPROTEIN(A); LIA; QUANTITATIVE ANALYSIS;
D O I
10.1002/jcla.1860080608
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
We had an opportunity to test a reagent kit for serum liporprotein (a) [Lp(a)] assay, which was based on the latex immunoturbidimetric assay method, and applied it to COBAS MIRA (Roche). Reproducibility of the assay procedure was 0.92-4.0% CV of three samples which contained 12.58-60.4 mg/dl of Lp(a). Minimum detectable concentration was 1.5-2.0 mg/dl. It was confirmed that interference of co-existing substances, i.e. hemoglobin, bilirubin, and intralipid, was negligible. No cross reactivity was seen with plasminogen. Correlation with a ELISA method was excellent, Frequency distribution of Lp(a) in healthy Japanese was similar to that found in white population. (C) 1994 Wiley-Liss, Inc.
引用
收藏
页码:385 / 390
页数:6
相关论文
共 50 条
  • [31] ENZYMATIC MEASUREMENT OF SERUM GLYCEROL USING A COBAS-BIO CENTRIFUGAL ANALYZER
    OU, CN
    FRAWLEY, VL
    CLINICAL BIOCHEMISTRY, 1985, 18 (01) : 37 - 39
  • [32] Homogenous enzyme immunoassay for cyclosporine in whole blood using the EMIT® 2000 cyclosporine specific assay with the COBAS MIRA-Plus analyzer
    Kimura, S
    Iyama, S
    Yamaguchi, Y
    Kanakura, Y
    JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2001, 15 (06) : 319 - 323
  • [33] Analytical validation of a point-of-care test and an automated immunoturbidimetric assay for the measurement of canine C-reactive protein in serum
    Covin, Marshal A.
    Gomez, Robynne R.
    Suchodolski, Jan S.
    Steiner, Jorg M.
    Lidbury, Jonathan A.
    CANADIAN JOURNAL OF VETERINARY RESEARCH-REVUE CANADIENNE DE RECHERCHE VETERINAIRE, 2021, 85 (04): : 285 - 292
  • [34] MEASUREMENT OF SERUM PYRIDOXAL USING AN AUTOMATED MICROBIOLOGICAL ASSAY
    DAVIS, RE
    CURNOW, DH
    PATHOLOGY, 1974, 6 (02) : 208 - 208
  • [35] Whole blood DNA extraction using the Roche COBAS® AmpliPrep® TNAI kit for use with the LightCycler Factor V Leiden assay
    Mendiola, J.
    Ducker, G.
    Ikier, D.
    Peterson, A.
    Rutledge, B.
    Reinartz, J.
    JOURNAL OF MOLECULAR DIAGNOSTICS, 2007, 9 (05): : 701 - 701
  • [36] AUTOMATED TURBIDIMETRIC ASSAY OF SERUM APOLIPOPROTEIN-A1 USING THE COBAS-BIO CENTRIFUGAL ANALYZER
    AUSTIN, GE
    MAZNICKI, E
    CLINICAL BIOCHEMISTRY, 1983, 16 (06) : 338 - 340
  • [37] EVALUATION OF ATLANTIC ANTIBODIES PREALBUMIN KIT METHOD ON THE COBAS FARA-II USING TURBIDIMETRIC AND NEPHELOMETRIC MEASUREMENT MODES
    ROSS, SJ
    KOCH, DD
    FELDBRUEGGE, DH
    CLINICAL CHEMISTRY, 1990, 36 (06) : 969 - 969
  • [38] A RAPID AND PRECISE METHOD FOR THE QUANTITATION OF ETHANOL IN SERUM USING THE EMIT SERUM ETHANOL ASSAY REAGENT AND A COBAS BIO CENTRIFUGAL ANALYZER
    DEGEL, F
    PAULUS, N
    JOURNAL OF CLINICAL CHEMISTRY AND CLINICAL BIOCHEMISTRY, 1988, 26 (05): : 351 - 353
  • [39] Evaluation of Analytical Measurement Ranges of Three Full Range C-Reactive Protein Tests Using Immunoturbidimetric Assay
    Lee, Young Jin
    Kim, Joong Man
    Park, Do-Sim
    KOREAN JOURNAL OF LABORATORY MEDICINE, 2010, 30 (01): : 9 - 16
  • [40] DETERMINATION OF FREE HEMOGLOBIN IN SERUM BY AN AUTOMATED-ASSAY USING 4-AMINOPHENAZONE AND THE COBAS BIO SYSTEM
    BAUER, K
    GABL, F
    JOURNAL OF CLINICAL CHEMISTRY AND CLINICAL BIOCHEMISTRY, 1981, 19 (08): : 610 - 610