MAGNETIC RESONANCE IMAGING;
FLOW;
ANGIOGRAPHY;
FAST IMAGING;
D O I:
10.1002/mrm.1910310219
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
We describe a time of flight subtraction method for cine MR angiography that provides nearly perfect suppression of background signal intensity with excellent flow contrast. The method consists of a preparation phase, during which the longitudinal magnetization of the target tissue is inverted on alternate acquisitions and the background tissue is presaturated, followed by a readout phase using a cine segmented turboFLASH sequence with a shared echo modification to improve temporal resolution. With appropriate alternation of the phases of the radiofrequency excitation pulses, there is cancellation of the background signal intensity but flow signal is optimized. By using a thick section (up to 25 mm), substantial portions of the vascular territory are encompassed in a single plane. This permits rapid, dynamic assessment of flow patterns in areas such as the circle of Willis, carotid bifurcation, or renal arteries. Applications of the method for bright and dark blood cine MR angiography are demonstrated.