Technical note: validation of cardiac output monitoring tools

被引:1
作者
Squara, P. [1 ]
Chiche, J. -D. [2 ]
机构
[1] Clin Ambroise Pare, Serv Reanimat CERIC, 27 Blvd Victor Hugo, F-92200 Neuilly, France
[2] Univ Cochin Port Royal, Ctr Hosp, Serv Reanimat Med, F-75013 Paris, France
来源
REANIMATION | 2011年 / 20卷 / 02期
关键词
Cardiac output; Monitoring; Bias; Precision; Variability; Repeatability; Least significant change;
D O I
10.1007/s13546-011-0224-4
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To date, tools for continuous cardiac output (CO) monitoring have been validated as if they were tools for snapshot measurements. Most authors compared variations in cardiac output between two time-points and used Bland & Altman representations to describe the agreement between these variations. The impact of repetitive measurements over time is not taken into consideration. This article proposes a conceptual framework for the validation of CO monitoring devices. The proposal consists of eight quality criteria: 1) accuracy (small bias); 2) linearity by comparing the regression line of the study and reference technology with the identity line; 3) precision (small random error of measurements); 4) least significant change; 5) stability (absence of time drift for the first four criteria); 6) measurement range where the first five criteria are fulfilled; 7) short response time; and 8) accurate amplitude response. Because tolerance is obviously admitted for each of these eight criteria, we propose to add as final criterion, the ability to detect significant CO directional changes. Other important issues in designing studies to validate CO monitoring tools are reviewed: choice of patient population to be studied, choice of the reference method, data acquisition method, data acceptability checking, data segmentation and final evaluation of reliability. Application of this framework will underline the importance of precision and time response for clinical acceptability of monitoring tools.
引用
收藏
页码:129 / 137
页数:9
相关论文
共 41 条
  • [21] Performance of noninvasive partial CO2 rebreathing cardiac output and continuous thermodilution cardiac output in patients undergoing aortic reconstruction surgery
    Kotake, Y
    Moriyama, K
    Innami, Y
    Shimizu, H
    Ueda, T
    Morisaki, H
    Takeda, J
    [J]. ANESTHESIOLOGY, 2003, 99 (02) : 283 - 288
  • [22] LeTulzo Y, 1996, J CLIN MONITOR, V12, P379
  • [23] Continuous cardiac output monitoring in humans by invasive and noninvasive peripheral blood pressure waveform analysis
    Lu, Zhenwei
    Mukkamala, Ramakrishna
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 2006, 101 (02) : 598 - 608
  • [24] Luchette FA, 2000, J INVEST SURG, V13, P147
  • [25] Comparison between Flotrac-Vigileo and Bioreactance, a totally noninvasive method for cardiac output monitoring
    Marque, Sophie
    Cariou, Alain
    Chiche, Jean-Daniel
    Squara, Pierre
    [J]. CRITICAL CARE, 2009, 13 (03)
  • [26] Noninvasive cardiac output by partial CO2 rebreathing after severe chest trauma
    Maxwell, RA
    Gibson, JB
    Slade, JB
    Fabian, TC
    Proctor, KG
    [J]. JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2001, 51 (05): : 849 - 853
  • [27] Validation of continuous thermodilution cardiac output in critically ill patients with analysis of systematic errors
    Medin, DL
    Brown, DT
    Wesley, R
    Cunnion, RE
    Ognibene, FP
    [J]. JOURNAL OF CRITICAL CARE, 1998, 13 (04) : 184 - 189
  • [28] A multicenter evaluation of a new continuous cardiac output pulmonary artery catheter system
    Mihm, FG
    Gettinger, A
    Hanson, CW
    Gilbert, HC
    Stover, EP
    Vender, JS
    Beerle, B
    Haddow, G
    [J]. CRITICAL CARE MEDICINE, 1998, 26 (08) : 1346 - 1350
  • [29] Passive leg raising
    Monnet, Xavier
    Teboul, Jean-Louis
    [J]. INTENSIVE CARE MEDICINE, 2008, 34 (04) : 659 - 663
  • [30] CARDIAC-OUTPUT MEASUREMENT BY THE THERMODILUTION METHOD - AN IN-VITRO TEST OF ACCURACY OF 3 COMMERCIALLY AVAILABLE AUTOMATIC CARDIAC-OUTPUT COMPUTERS
    RUBINI, A
    DELMONTE, D
    CATENA, V
    ATTAR, I
    CESARO, M
    SORANZO, D
    RATTAZZI, G
    ALATI, GL
    [J]. INTENSIVE CARE MEDICINE, 1995, 21 (02) : 154 - 158