Dual energy CT angiography of the carotid arteries: quality, bone subtraction, and radiation dosage using tube voltage 80/140kV versus 100/140kV

被引:1
|
作者
Santos Armentia, E. [1 ]
Tardaguila de la Fuente, G. [1 ]
Castellon Plaza, D. [1 ]
Delgado Sanchez-Gracian, C. [1 ]
Prada Gonzalez, R. [1 ]
Fernandez Fernandez, L. [1 ]
Tardaguila Montero, F. [1 ]
机构
[1] Hosp Povisa, Serv Radiodiagnost, Vigo, Pontevedra, Spain
来源
RADIOLOGIA | 2014年 / 56卷 / 03期
关键词
Computed tomography; Dual sourceCT; Angiography CT angiography; Supraaortic trunks; Radiation dose; Image quality; Bone removal;
D O I
10.1016/j.rx.2012.02.003
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To study the differences in vascular image quality, bone subtraction, and dose of radiation of dual energy CT angiography of the supraaortic trunks using different tube voltages. Material and methods: We reviewed the CT angiograms of the supraaortic trunks in 46 patients acquired with a 128-slice dual source CT scanner using two voltage protocols (80/140 kV and 100/140 kV). The "head bone removal" tool was used for postprocessing. We divided the arteries into 15 segments. In each segment, we evaluated the image quality of the vessels and the effectiveness of bone removal in multiplanar reconstructions (MPR) and in maximum intensity projections (MIP) with each protocol, analyzing the trabecular and cortical bones separately. We also evaluated the dose of radiation received. Results: Of the 46 patients, 13 were studied using 80/140 kV and 33 with 100/140 kV. There were no significant differences between the two groups in age or sex. Image quality in four segments was better in the group examined with 100/140 kV. Cortical bone removal in MPR and MIP and trabecular bone removal in MIP were also better in the group examined with 100/140 kV. The dose of radiation received was significantly higher in the group examined with 100/140 kV (1.16 mSv with 80/140 kV vs. 1.59 mSv with 100/140 kV). Conclusion: Using 100/140 kV increases the dose of radiation but improves the quality of the study of arterial segments and bone subtraction. (C) 2011 SERAM. Published by Elsevier Espana, S. L. All rights reserved.
引用
收藏
页码:241 / 246
页数:6
相关论文
共 3 条
  • [1] The effect of tube voltage combination on image artefact and radiation dose in dual-source dual-energy CT: comparison between conventional 80/140 kV and 80/150 kV plus tin filter for gout protocol
    Jeon, Ji Young
    Lee, Sheen-Woo
    Jeong, Yu Mi
    Baek, Han Joo
    EUROPEAN RADIOLOGY, 2019, 29 (03) : 1248 - 1257
  • [2] The use of 80 kV versus 100 kV in pulmonary CT angiography: An evaluation of the impact on radiation dose and image quality on two CT scanners
    Rusandu, A.
    Odegard, A.
    Engh, G. C.
    Olerud, H. M.
    RADIOGRAPHY, 2019, 25 (01) : 58 - 64
  • [3] Subtraction CT angiography in head and neck with low radiation and contrast dose dual-energy spectral CT using rapid kV-switching technique
    Ma, Guangming
    Yu, Yong
    Duan, Haifeng
    Dou, Yuequn
    Jia, Yongjun
    Zhang, Xirong
    Yang, Chuangbo
    Chen, Xiaoxia
    Han, Dong
    Guo, Changyi
    He, Taiping
    BRITISH JOURNAL OF RADIOLOGY, 2018, 91 (1086)