PCO2 CONTROL BY FRESH GAS FLOW DURING CONTROLLED VENTILATION WITH A SEMI-OPEN CIRCUIT

被引:27
作者
BARAKA, A
机构
[1] Department of Anesthesiology, American University of Beirut, Beirut
关键词
D O I
10.1093/bja/41.6.527
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Summary: Investigation was carried out on ten adult patients undergoing abdominal surgery under nitrous oxide, oxygen and alcuronium anaesthesia, using the Mapleson D semi-open system. Ventilation was controlled throughout at a constant ventilation volume of 16 l./min. Pco2 levels varied according to the fresh gas flow. The mean Pco2 was 34.7 (SD 5.7) mm Hg at a fresh gas flow of 8 l./min, 44.2 (SD 6.3) mm Hg at a flow of 4 l./min, 60.2 (SD 9.9) mm Hg at a flow of 2 l./min, and 88 (SD 17.6) mm Hg at a flow of 1 l./min. At normal body temperature, normocarbia will be achieved by a fresh gas flow of about 5 l./min, which is approximately equal to the alveolar ventilation volume of the average adult. © 1969 John Sherratt and Son Ltd.
引用
收藏
页码:527 / &
相关论文
共 8 条
[1]   REBREATHING IN A DOUBLE T-PIECE SYSTEM [J].
BARAKA, A ;
BRANDSTATER, B ;
MUALLEM, M ;
SERAPHIM, C .
BRITISH JOURNAL OF ANAESTHESIA, 1969, 41 (01) :47-+
[2]  
BROOKS D, 1959, LANCET, V1, P227
[3]  
GESSEL R, 1930, AM J PHYSIOL, V94, P402
[4]   FRESH GAS ECONOMICS OF MAGILL CIRCUIT [J].
KAIN, ML ;
NUNN, JF .
ANESTHESIOLOGY, 1968, 29 (05) :964-+
[5]  
MAPLESON W W, 1954, Br J Anaesth, V26, P323, DOI 10.1093/bja/26.5.323
[7]   INSPIRED GAS REBREATHING + APPARATUS DEADSPACE [J].
NUNN, JF ;
NEWMAN, HC .
BRITISH JOURNAL OF ANAESTHESIA, 1964, 36 (01) :5-+
[8]   REBREATHING DURING CONTROLLED RESPIRATION WITH VARIOUS SEMICLOSED ANAESTHETIC SYSTEMS [J].
WATERS, DJ ;
MAPLESON, WW .
BRITISH JOURNAL OF ANAESTHESIA, 1961, 33 (08) :374-&