RAPID MR MEASUREMENT OF CONTRAST-MEDIUM DILUTION KINETICS (GADOLINIUM-DTPA) IN A FLOW PHANTOM

被引:9
作者
BOCK, JC [1 ]
SANDER, B [1 ]
FRANK, J [1 ]
SCHORNER, W [1 ]
机构
[1] FREE UNIV BERLIN,RUDOLF VIRCHOW KLIN,CHIRURG KLIN & POLIKLIN,W-1000 BERLIN 19,GERMANY
来源
FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NEUEN BILDGEBENDEN VERFAHREN | 1991年 / 155卷 / 03期
关键词
INDICATOR DILUTION METHOD; PARAMAGNETIC CONTRAST MEDIA; MAGNETIC RESONANCE IMAGING; GRADIENT ECHO (FLASH);
D O I
10.1055/s-2008-1033259
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We studied first-pass MRI-contrast dilution to compute flow and volume of distribution in a realistic flow phantom. Pulsatile flow was provided by a one-chamber artificial heart. Physiological stroke volume, rate, pressure, and flow were adjustable. An elastic tube with dimensions similar to that of the human aorta was imaged at a rate of 2.4 Hz. After contrast injection, an initial increase in signal intensity was followed by a decrease. Signal-intensity-time plots demonstrated slightly skewed curves as expected from dispersion theory. After calibration at different gadolinium-DTPA concentrations, signal intensities were converted into true gadolinium concentrations, and flow was calculated from the concentration-time curves. Flow was varied between 2.5 and 10.0 l/min and a significant correlation was found between the MRI-estimate and true flow. Volume of distribution between injection and detection site was reliably estimated. This study demonstrates rapid 2-D imaging of a paramagnetic contrast bolus in a realistic flow phantom. Reliable estimates of flow and volume are obtained.
引用
收藏
页码:267 / 271
页数:5
相关论文
共 11 条
[1]   1ST-PASS CARDIAC PERFUSION - EVALUATION WITH ULTRAFAST MR IMAGING [J].
ATKINSON, DJ ;
BURSTEIN, D ;
EDELMAN, RR .
RADIOLOGY, 1990, 174 (03) :757-762
[2]   THERMAL RECOVERY AFTER PASSAGE OF THE PULMONARY CIRCULATION ASSESSED BY DECONVOLUTION [J].
BOCK, J ;
DEUFLHARD, P ;
HOEFT, A ;
KORB, H ;
WOLPERS, HG ;
STEINMANN, J ;
HELLIGE, G .
JOURNAL OF APPLIED PHYSIOLOGY, 1988, 64 (03) :1210-1216
[3]   NO PULMONARY-EDEMA OR CONGESTION AFTER CENTRAL VENOUS INJECTION OF CONVENTIONAL AND NEWER CONTRAST-MEDIA IN DOGS [J].
BOCK, J ;
HEILBRON, DC ;
HOEFT, A ;
KORB, H ;
HELLIGE, G .
INVESTIGATIVE RADIOLOGY, 1988, 23 (11) :836-841
[4]  
BOCK JC, 1990, INVEST RADIOL, V25, P938
[5]  
BOCK JC, 1990, 19 ANN M SOC MAGN RE
[6]   INDICATOR DILUTION TIME-ACTIVITY CURVES DEMONSTRATED BY RAPID MAGNETIC-RESONANCE IMAGING TECHNIQUES AND PARAMAGNETIC CONTRAST AGENT [J].
DAVIS, PL ;
WOLF, GL ;
GILLEN, JS .
INVESTIGATIVE RADIOLOGY, 1989, 24 (05) :400-406
[7]  
FRANK J, 1979, BIOMED TECHNIK, V24, P307
[8]  
FRESTIN GP, 1988, DIAGN IMAG INT S, V40
[9]   DYNAMIC IMAGING WITH LANTHANIDE CHELATES IN NORMAL BRAIN - CONTRAST DUE TO MAGNETIC-SUSCEPTIBILITY EFFECTS [J].
VILLRINGER, A ;
ROSEN, BR ;
BELLIVEAU, JW ;
ACKERMAN, JL ;
LAUFFER, RB ;
BUXTON, RB ;
CHAO, YS ;
WEDEEN, VJ ;
BRADY, TJ .
MAGNETIC RESONANCE IN MEDICINE, 1988, 6 (02) :164-174
[10]  
WEINMANN HJ, 1989, ENHANCED MAGNETIC RE, P74