TECHNICAL NOTE - BONE SPECT OF THE WRIST

被引:7
作者
DUBOWITZ, DJ [1 ]
MILES, KA [1 ]
机构
[1] ADDENBROOKES NATL HLTH SERV TRUST,DEPT NUCL MED,CAMBRIDGE CB2 2QQ,CAMBS,ENGLAND
关键词
D O I
10.1259/0007-1285-67-801-890
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A technique for single photon emission computed tomography (SPECT) during bone scintigraphy of the wrist is described. The technique has been developed as a means to improve localization of abnormalities observed on Tc-99(m)-MDP planar bone scintigraphy. A wrist phantom and 15 wrists in 13 symptomatic patients were examined. Good spatial detail was achieved, with the pisiform readily identifiable as a separate bone in six wrists. In seven cases the tomograms permitted better localization of planar image abnormalities and in four cases additional lesions were visible which were missed on planar scintigraphy. This technique can be applied to all patients with suspected carpal injury and allows better anatomical localization and greater sensitivity than planar scintigraphy alone.
引用
收藏
页码:890 / 893
页数:4
相关论文
共 5 条
[1]  
COLLIER BD, 1987, BONE SCANNING CLIN P, P175
[2]   REGISTRATION AND DISPLAY OF THE COMBINED BONE-SCAN AND RADIOGRAPH IN THE DIAGNOSIS AND MANAGEMENT OF WRIST INJURIES [J].
HAWKES, DJ ;
ROBINSON, L ;
CROSSMAN, JE ;
SAYMAN, HB ;
MISTRY, R ;
MAISEY, MN ;
SPENCER, JD .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1991, 18 (09) :752-756
[3]  
PATEL N, 1993, CLIN NUCL MED, V17, P449
[4]  
VANBUUL MMT, 1993, EUR J NUCL MED, V20, P159
[5]  
VANBUUL MMT, 1993, INJURY, V24, P257