A ROTATING PET SCANNER USING BGO BLOCK DETECTORS - DESIGN, PERFORMANCE AND APPLICATIONS

被引:0
作者
TOWNSEND, DW
WENSVEEN, M
BYARS, LG
GEISSBUHLER, A
TOCHONDANGUY, HJ
CHRISTIN, A
DEFRISE, M
BAILEY, DL
GROOTOONK, S
DONATH, A
NUTT, R
机构
[1] UNIV GENEVA,HOP CANTONAL,DIV NUCL MED,CH-1211 GENEVA 4,SWITZERLAND
[2] VRIJE UNIV BRUSSELS,DIV NUCL MED,B-1050 BRUSSELS,BELGIUM
[3] CTI SIEMENS PET SYST,KNOXVILLE,TN
[4] ROYAL PRINCE ALFRED HOSP,DIV NUCL MED,SYDNEY,AUSTRALIA
[5] HAMMERSMITH HOSP,MRC,CYCLOTRON UNIT,LONDON W12 0HS,ENGLAND
关键词
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Recent advances in fully three-dimensional reconstruction for multi-ring PET scanners have led us to explore the potential of a prototype scanner based on the rotation of two opposing arrays of BGO block detectors. The prototype contains only one-third of the number of detectors in the equivalent full ring scanner, resulting in reduced cost. With a lower energy threshold at 250 keV, the absolute efficiency of the scanner is 0.5% and the scatter fraction is 35% for a 20-cm cylinder. Transaxial and axial spatial resolution is about 6 mm. The maximum noise equivalent count rate estimated for a 15-cm diameter cylinder is 36,000 cps at a concentration of 26 kBq/ml. The minimum scan time for a F-18-fluoro-2-deoxyglucose (FDG) brain study is 55 sec. The camera has been validated for clinical applications using both FDG and Rb-82.
引用
收藏
页码:1367 / 1376
页数:10
相关论文
共 24 条
  • [1] BAILEY D L, 1989, European Journal of Nuclear Medicine, V15, P452
  • [2] NOISE EQUIVALENT COUNT MEASUREMENTS IN A NEURO-PET SCANNER WITH RETRACTABLE SEPTA
    BAILEY, DL
    JONES, T
    SPINKS, TJ
    GILARDI, MC
    TOWNSEND, DW
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1991, 10 (03) : 256 - 260
  • [3] BROWNELL GL, 1977, RECONSTRUCTION TOMOG, P293
  • [4] A MULTICRYSTAL 2-DIMENSIONAL BGO DETECTOR SYSTEM FOR POSITRON EMISSION TOMOGRAPHY
    CASEY, ME
    NUTT, R
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1986, 33 (01) : 460 - 463
  • [5] FULLY 3-DIMENSIONAL POSITRON EMISSION TOMOGRAPHY
    COLSHER, JG
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1980, 25 (01) : 103 - 115
  • [6] IMPLEMENTATION OF 3-DIMENSIONAL IMAGE-RECONSTRUCTION FOR MULTIRING POSITRON TOMOGRAPHS
    DEFRISE, M
    TOWNSEND, D
    GEISSBUHLER, A
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (10) : 1361 - 1372
  • [7] A NEW 3-DIMENSIONAL RECONSTRUCTION METHOD FOR POSITRON CAMERAS USING PLANE DETECTORS
    DEFRISE, M
    KUIJK, S
    DECONINCK, F
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1988, 33 (01) : 43 - 51
  • [8] DEFRISE M, 1992, IEEE MED IM C, V3, P1919
  • [9] TOMOGRAPHIC IMAGING OF THE HUMAN THYROID USING I-124
    FREY, P
    TOWNSEND, D
    FLATTET, A
    DEGAUTARD, R
    WIDGREN, S
    JEAVONS, A
    CHRISTIN, A
    SMITH, A
    LONG, A
    DONATH, A
    [J]. JOURNAL OF CLINICAL ENDOCRINOLOGY & METABOLISM, 1986, 63 (04) : 918 - 927
  • [10] KINAHAN PE, 1990, IEEE T NUCL SCI, V3, P964