TIDAL VOLUME MEASUREMENTS IN NEWBORNS USING RESPIRATORY INDUCTIVE PLETHYSMOGRAPHY

被引:60
作者
ADAMS, JA
ZABALETA, IA
STROH, D
JOHNSON, P
SACKNER, MA
机构
[1] MT SINAI MED CTR,DIV PULM DIS,MIAMI BEACH,FL 33140
[2] FLORIDA INT UNIV,DEPT STAT,MIAMI,FL 33199
来源
AMERICAN REVIEW OF RESPIRATORY DISEASE | 1993年 / 148卷 / 03期
关键词
D O I
10.1164/ajrccm/148.3.585
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
Respiratory inductive plethysmography (RIP) is a well-accepted noninvasive technology for monitoring breathing patterns in adults. Prior attempts to calibrate this device in babies have been fraught with technical difficulties, thereby limiting applications in this population. Recently, a new method, qualitative diagnostic calibration (QDC), has been shown to provide accurate calibration of tidal volume in adults. The QDC method is based upon principles of the isovolume maneuver and carried out during natural breathing without specialized respiratory maneuvers or postural changes. We calibrated RIP with QDC in the supine posture and compared tidal volume (VT) measured with RIP to VT by a face mask-pneumotachograph (PNT) in 21 healthy full-term newborns in supine and prone postures. Eleven of the babies were calibrated during active sleep and 10 in quiet sleep. The mean VT in the supine and prone postures were 19 and 25 ml, respectively. In the supine and prone postures, weighted mean difference between RIP (VT) and PNT (VT) and 95% confidence intervals were -0.05 ml (-0.27, 0.18) and -0.32 ml (-0.08, 0.55), respectively. There was no difference in the accuracy of RIP relative to PNT calibrated during active sleep when thoracoabdominal incoordination was present or quiet sleep when it was not in either the supine or the prone postures. Therefore, in full-term infants, RIP calibrated with QDC solely in the supine posture provides clinically acceptable measurements of VT in both supine and prone postures.
引用
收藏
页码:585 / 588
页数:4
相关论文
共 17 条
  • [1] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [2] CHADHA TS, 1982, AM REV RESPIR DIS, V125, P644
  • [3] DOLFIN T, 1982, AM REV RESPIR DIS, V126, P577
  • [4] DOLFIN T, 1983, AM REV RESPIR DIS, V128, P977
  • [5] DRAPER NR, 1981, APPL REGRESSION ANAL, P108
  • [6] DUFFTY P, 1981, AM REV RESPIR DIS, V123, P542
  • [7] FAMILIES OF LINES - RANDOM EFFECTS IN LINEAR-REGRESSION ANALYSIS
    FELDMAN, HA
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 1988, 64 (04) : 1721 - 1732
  • [8] THE SLEEP HYPOPNEA SYNDROME
    GOULD, GA
    WHYTE, KF
    RHIND, GB
    AIRLIE, MAA
    CATTERALL, JR
    SHAPIRO, CM
    DOUGLAS, NJ
    [J]. AMERICAN REVIEW OF RESPIRATORY DISEASE, 1988, 137 (04): : 895 - 898
  • [9] MEASUREMENT OF SEPARATE VOLUME CHANGES OF RIB CAGE AND ABDOMEN DURING BREATHING
    KONNO, K
    MEAD, J
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 1967, 22 (03) : 407 - &
  • [10] COMPARISON OF CALIBRATION METHODS FOR RESPIRATORY INDUCTIVE PLETHYSMOGRAPHY IN INFANTS
    REVOW, MD
    ENGLAND, SJ
    STOGRYN, HAF
    WILKES, DL
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 1987, 63 (05) : 1853 - 1861