Four-dimensional Plan Optimization for the Treatment of Lung Tumors Using Pencil-beam Scanning Proton Radiotherapy

被引:19
作者
Cummings, David [1 ]
Tang, Shikui [2 ]
Ichter, William [1 ]
Wang, Peng [1 ]
Sturgeon, Jared D. [3 ]
Lee, Andrew K. [3 ]
Chang, Chang [2 ]
机构
[1] Texas Ctr Proton Therapy, Phys, Irving, TX USA
[2] Texas Ctr Proton Therapy, Med Phys, Irving, TX 75063 USA
[3] Texas Ctr Proton Therapy, Radiat Oncol, Irving, TX USA
关键词
4d robust optimization; robust optimization; proton pbs treatment; interplay effect;
D O I
10.7759/cureus.3192
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Purpose This study aimed to evaluate the effectiveness of four-dimensional (4D) robust optimization for proton pencil-beam scanning (PBS) treatment of lung tumors. Patients and methods In seven patients with lung cancer, proton beam therapy was planned using 4D robust optimization over 4D computed tomography (CT) data sets. The gross target volume (GTV) was contoured based on individual breathing phases, and a 5-mm expansion was used to generate the clinical target volume (cry) for each phase. The 4D optimization was conducted directly on the 4D CT data set. The robust optimization settings included a CT Hounsfield unit (HU) uncertainty of 4% and a setup uncertainty of 5 mm to obtain the CTV. Additional target dose objectives such as those for the internal target volume (ITV) as well as the organ at risk (OAR) dose requirements were placed on the average CT. For comparison, three-dimensional (3D) robust optimization was also performed on the average CT. An additional verification 4D CT was performed to verify plan robustness against inter-fractional variations. Results Target coverages were generally higher for 4D optimized plans. The difference was most pronounced for ITV V70Gy when evaluating individual breathing phases. The 4D optimized plans were shown to he able to maintain the ITV coverage at full prescription, while 3D optimized plans could not. More importantly, this difference in ITV V70Gy between the 4D and 3D optimized plans was also consistently observed when evaluating the verification 4D CT, indicating that the 4D optimized plans were more robust against inter fractional variations. Less difference was seen between the 4D and 3D optimized plans in the lungs criteria: V5Gy and V20Gy. Conclusion The proton PBS treatment plans optimized directly on the 4D CT were shown to be more robust when compared to those optimized on a regular 3D CT. Robust 4D optimization can improve the target coverage for the proton PBS lung treatments.
引用
收藏
页数:9
相关论文
共 30 条
[21]   Monte Carlo simulations with time-dependent geometries to investigate effects of organ motion with high temporal resolution [J].
Paganetti, H ;
Jiang, HY ;
Adams, JA ;
Chen, GT ;
Rietzel, E .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (03) :942-950
[22]   Upgrade and benchmarking of a 4D treatment planning system for scanned ion beam therapy [J].
Richter, D. ;
Schwarzkopf, A. ;
Trautmann, J. ;
Kraemer, M. ;
Durante, M. ;
Jaekel, O. ;
Bert, C. .
MEDICAL PHYSICS, 2013, 40 (05)
[23]   Maximum-intensity volumes for fast contouring of lung tumors including respiratory motion in 4DCT planning [J].
Rietzel, Eike ;
Liu, Arthur K. ;
Chen, George T. Y. ;
Choi, Noah C. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 71 (04) :1245-1252
[24]   Breathing interplay effects during proton beam scanning: simulation and statistical analysis [J].
Seco, Joao ;
Robertson, Daniel ;
Trofimov, Alexei ;
Paganetti, Harald .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (14) :N283-N294
[25]   Cancer Statistics, 2017 [J].
Siegel, Rebecca L. ;
Miller, Kimberly D. ;
Jemal, Ahmedin .
CA-A CANCER JOURNAL FOR CLINICIANS, 2017, 67 (01) :7-30
[26]   Potentials of robust intensity modulated scanning proton plans for locally advanced lung cancer in comparison to intensity modulated photon plans [J].
Stuschke, Martin ;
Kaiser, Andreas ;
Poettgen, Christoph ;
Luebcke, Wolfgang ;
Farr, Jonathan .
RADIOTHERAPY AND ONCOLOGY, 2012, 104 (01) :45-51
[27]   Use of maximum intensity projections (MIP) for target volume generation in 4DCT scans for lung cancer [J].
Underberg, RWM ;
Lagerwaard, FJ ;
Slotman, BJ ;
Cuijpers, JP ;
Senan, S .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (01) :253-260
[28]   Evaluation and comparison of New 4DCT based strategies for proton treatment planning for lung tumors [J].
Wang, Ning ;
Patyal, Baldev ;
Ghebremedhin, Abiel ;
Bush, David .
RADIATION ONCOLOGY, 2013, 8
[29]   Clinical examination of proton pencil beam scanning on a moving anthropomorphic lung phantom [J].
Wang, Peng ;
Tang, Shikui ;
Taylor, Paige A. ;
Cummings, David E. ;
Janson, Martin ;
Traneus, Erik ;
Sturgeon, Jared D. ;
Lee, Andrew K. ;
Chang, Chang .
MEDICAL DOSIMETRY, 2019, 44 (02) :122-129
[30]   Prospective Study of Patient-Reported Symptom Burden in Patients With Non-Small-Cell Lung Cancer Undergoing Proton or Photon Chemoradiation Therapy [J].
Wang, Xin Shelley ;
Shi, Qiuling ;
Williams, Loretta A. ;
Komaki, Ritsuko ;
Gomez, Daniel R. ;
Lin, Steven H. ;
Chang, Joe Y. ;
O'Reilly, Michael S. ;
Bokhari, Raza H. ;
Cox, James D. ;
Mohan, Radhe ;
Cleeland, Charles S. ;
Liao, Zhongxing .
JOURNAL OF PAIN AND SYMPTOM MANAGEMENT, 2016, 51 (05) :832-838