Four-dimensional Plan Optimization for the Treatment of Lung Tumors Using Pencil-beam Scanning Proton Radiotherapy

被引:19
作者
Cummings, David [1 ]
Tang, Shikui [2 ]
Ichter, William [1 ]
Wang, Peng [1 ]
Sturgeon, Jared D. [3 ]
Lee, Andrew K. [3 ]
Chang, Chang [2 ]
机构
[1] Texas Ctr Proton Therapy, Phys, Irving, TX USA
[2] Texas Ctr Proton Therapy, Med Phys, Irving, TX 75063 USA
[3] Texas Ctr Proton Therapy, Radiat Oncol, Irving, TX USA
关键词
4d robust optimization; robust optimization; proton pbs treatment; interplay effect;
D O I
10.7759/cureus.3192
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Purpose This study aimed to evaluate the effectiveness of four-dimensional (4D) robust optimization for proton pencil-beam scanning (PBS) treatment of lung tumors. Patients and methods In seven patients with lung cancer, proton beam therapy was planned using 4D robust optimization over 4D computed tomography (CT) data sets. The gross target volume (GTV) was contoured based on individual breathing phases, and a 5-mm expansion was used to generate the clinical target volume (cry) for each phase. The 4D optimization was conducted directly on the 4D CT data set. The robust optimization settings included a CT Hounsfield unit (HU) uncertainty of 4% and a setup uncertainty of 5 mm to obtain the CTV. Additional target dose objectives such as those for the internal target volume (ITV) as well as the organ at risk (OAR) dose requirements were placed on the average CT. For comparison, three-dimensional (3D) robust optimization was also performed on the average CT. An additional verification 4D CT was performed to verify plan robustness against inter-fractional variations. Results Target coverages were generally higher for 4D optimized plans. The difference was most pronounced for ITV V70Gy when evaluating individual breathing phases. The 4D optimized plans were shown to he able to maintain the ITV coverage at full prescription, while 3D optimized plans could not. More importantly, this difference in ITV V70Gy between the 4D and 3D optimized plans was also consistently observed when evaluating the verification 4D CT, indicating that the 4D optimized plans were more robust against inter fractional variations. Less difference was seen between the 4D and 3D optimized plans in the lungs criteria: V5Gy and V20Gy. Conclusion The proton PBS treatment plans optimized directly on the 4D CT were shown to be more robust when compared to those optimized on a regular 3D CT. Robust 4D optimization can improve the target coverage for the proton PBS lung treatments.
引用
收藏
页数:9
相关论文
共 30 条
[1]   Effects of intra-fraction motion on IMRT dose delivery: statistical analysis and simulation [J].
Bortfeld, T ;
Jokivarsi, K ;
Goitein, M ;
Kung, J ;
Jiang, SB .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (13) :2203-2220
[2]   Treatment Parameters Optimization to Compensate for Interfractional Anatomy Variability and Intrafractional Tumor Motion [J].
Brevet, Romain ;
Richter, Daniel ;
Graeff, Christian ;
Durante, Marco ;
Bert, Christoph .
FRONTIERS IN ONCOLOGY, 2015, 5
[3]   Long-term outcome of phase I/II prospective study of dose-escalated proton therapy for early-stage non-small cell lung cancer [J].
Chang, Joe Y. ;
Zhang, Wencheng ;
Komaki, Ritsuko ;
Choi, Noah C. ;
Chan, Shen ;
Gomez, Daniel ;
O'Reilly, Michael ;
Jeter, Melenda ;
Gillin, Michael ;
Zhu, Xiaorong ;
Zhang, Xiaodong ;
Mohan, Radhe ;
Swisher, Stephen ;
Hahn, Stephen ;
Cox, James D. .
RADIOTHERAPY AND ONCOLOGY, 2017, 122 (02) :274-280
[4]   Consensus Statement on Proton Therapy in Early-Stage and Locally Advanced Non-Small Cell Lung Cancer [J].
Chang, Joe Y. ;
Jabbour, Salma K. ;
De Ruysscher, Dirk ;
Schild, Steven E. ;
Simone, Charles B., II ;
Rengan, Ramesh ;
Feigenberg, Steven ;
Khan, Atif J. ;
Choi, Noah C. ;
Bradley, Jeffrey D. ;
Zhu, Xiaorong R. ;
Lomax, Antony J. ;
Hoppe, Bradford S. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 95 (01) :505-516
[5]   Interplay effects in proton scanning for lung: a 4D Monte Carlo study assessing the impact of tumor and beam delivery parameters [J].
Dowdell, S. ;
Grassberger, C. ;
Sharp, G. C. ;
Paganetti, H. .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (12) :4137-4156
[6]   4D optimization of scanned ion beam tracking therapy for moving tumors [J].
Eley, John Gordon ;
Newhauser, Wayne David ;
Luechtenborg, Robert ;
Graeff, Christian ;
Bert, Christoph .
PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (13) :3431-3452
[7]   Feasibility of Pencil Beam Scanned Intensity Modulated Proton Therapy in Breath-hold for Locally Advanced Non-Small Cell Lung Cancer [J].
Gorgisyan, Jenny ;
af Rosenschold, Per Munck ;
Perrin, Rosalind ;
Persson, Gitte F. ;
Josipovic, Mirjana ;
Belosi, Maria Francesca ;
Engelholm, Svend Aage ;
Weber, Damien C. ;
Lomax, Antony J. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2017, 99 (05) :1121-1128
[8]   Robustness of 4D-optimized scanned carbon ion beam therapy against interfractional changes in lung cancer [J].
Graeff, Christian .
RADIOTHERAPY AND ONCOLOGY, 2017, 122 (03) :387-392
[9]   Motion mitigation for lung cancer patients treated with active scanning proton therapy [J].
Grassberger, Clemens ;
Dowdell, Stephen ;
Sharp, Greg ;
Paganetti, Harald .
MEDICAL PHYSICS, 2015, 42 (05) :2462-2469
[10]   Quantification of Proton Dose Calculation Accuracy in the Lung [J].
Grassberger, Clemens ;
Daartz, Juliane ;
Dowdell, Stephen ;
Ruggieri, Thomas ;
Sharp, Greg ;
Paganetti, Harald .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 89 (02) :424-430