SPATIAL AND TEMPORAL RESPONSE CHARACTERISTICS OF IONIZATION CHAMBERS USED IN DIAGNOSTIC-RADIOLOGY FOR EXPOSURE MEASUREMENTS AND QUALITY-CONTROL

被引:5
作者
KOFLER, JM
GRAY, JE
DALY, TR
机构
[1] Department of Diagnostic Radiology, Mayo Clinic and Foundation, Rochester, MN
来源
HEALTH PHYSICS | 1994年 / 67卷 / 06期
关键词
DIAGNOSTIC RADIOLOGY; IONIZATION CHAMBERS; OPERATIONAL TOPICS; QUALITY CONTROL;
D O I
10.1097/00004032-199412000-00010
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The temporal and spatial (partial volume) response characteristics of ionization chambers used for measuring radiation exposures in x-ray quality control (QC) programs were evaluated. Five ionization chambers were evaluated using a pencil beam scanning x-ray source and a conventional radiographic system. The spatial response was determined by recording the exposure during scanning of the pencil beam or in increments using a slitted lead sheet on a conventional x-ray system. The temporal response was determined by recording x-ray wave forms using the different ionization chambers. The effects of partial volume irradiation of the chambers makes them unsuitable for use under these conditions, except those designed for exposure measurements in computed tomography. The temporal response of many chambers resembles that of a typical resistive-capacitive circuit, making them unsuitable for exposure time measurements or x-ray wave form evaluation. The appropriate ionization chamber must be selected for exposure measurements and quality control in diagnostic radiology.
引用
收藏
页码:661 / 667
页数:7
相关论文
共 4 条
[1]  
Attix FH., 1986, INTRO RADIOLOGICAL P, P358
[2]  
JOHNS HE, 1983, PHYSICS RADIOLOGY
[3]  
Knoll GF, 1979, RAD DETECTION MEASUR
[4]  
SUNDE PB, 1992, RADIAT PROT DOSIM, V43, P183