COMPARATIVE-STUDY OF 2 LASER DOPPLER BLOOD FLOWMETERS

被引:34
作者
BARNETT, NJ [1 ]
DOUGHERTY, G [1 ]
PETTINGER, SJ [1 ]
机构
[1] ICI PHARMACEUT PLC,MACCLESFIELD SK10 4TF,CHESHIRE,ENGLAND
关键词
D O I
10.3109/03091909009015431
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A new infrared laser Doppler blood flow instument (Moor MBF3D) was evaluated using an in vitro model allowing measurements over a range of flow velocities and concentrations. The responses correlated well (r = 0.96, p < 0.01) with those obtained simultaneously using a Perimed PF3 laser Doppler instrument. The different processing bandwidths of the instruments were investigated and the wideband mode of operation is recommended for flow measurements where there may be fast moving red blood cells (rbcs). The infrared instrument is capable of dual-channel operation, and the two channels are shown to respond almost identically for similar changes in blood flow through the in vitro model (r = 0.999, p < 0.01). The main advantage of the dual-channel instrument is that continuous measurements may be made simultaneously at two different skin sites allowing dynamic flow responses to be compared.
引用
收藏
页码:243 / 249
页数:7