CONTRAST-DETAIL-DOSE ANALYSIS OF 6 DIFFERENT COMPUTED TOMOGRAPHIC SCANNERS

被引:61
作者
COHEN, G
机构
[1] ALLEGHENY GEN HOSP,RADIOL PHYS SECT,PITTSBURGH,PA 15212
[2] ALLEGHENY GEN HOSP,DIV DIAGNOST RADIOL,PITTSBURGH,PA 15212
关键词
Computed tomography; Image quality; Radiation dose;
D O I
10.1097/00004728-197904000-00010
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Low-contrast visibility of various rotate-translate, rotating detector, and stationary detector computed tomography (CT) scanners is measured at contrasts ranging from 0.3 to 3% using a phantom based on the partial volume of water and polystyrene. The results are plotted on log-log curves, as a function of entrance dose, to provide a family of contrast (C)-detail (d) —dose (D) visibility curves. The transition region, ~ 1 to 10% contrast, depends on modulation transfer function (MTF), contrast, and dose (noise), while the noise region, < 1% contrast, is independent of MTF. The slopes of the curves in the noise region are consistent with a value of -1, as opposed to a theoretically expected value of -3/2 for ideal CT noise. Relative dose efficiency of various scanners is examined by plotting CdD1/2versus contrast for both peak dose (Dp) and summation dose (DƩ). The rotating detector systems are shown to be, on the average (at 0.5% contrast), 3.4 (Dp) and 4 (DƩ) times more dose efficient than the stationary array system examined, and 2.4 (Dp) and 3.2 (DƩ) times more efficient than the rotate-translate units examined. © 1979 Raven Press, New York.
引用
收藏
页码:197 / 203
页数:7
相关论文
共 5 条
[1]   USE OF CONTRAST-DETAIL-DOSE EVALUATION OF IMAGE QUALITY IN A COMPUTED TOMOGRAPHIC SCANNER [J].
COHEN, G ;
DIBIANCA, FA .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1979, 3 (02) :189-195
[2]  
COULAM G, COMMUNICATION
[3]  
MASSE F, UNPUBLISHED
[4]  
WAGNER RF, 1978, MED PHYS, V5, P342
[5]   IMAGE QUALITY IN CRANIAL COMPUTED TOMOGRAPHY [J].
ZATZ, LM .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1978, 2 (03) :336-346