This study reports on the soft-tissue response to titanium fiber mesh percutaneous implants. The implants were inserted in the backs of 15 rabbits. The surgical procedure was performed in two steps: implantation of the subcutaneous part of the implants and, after 3 months, fixation of the percutaneous part into the subcutaneous part. Two and 6 months after insertion the animals were sacrificed and the implants, with their surrounding tissues, were processed histologically. Light microscopic analysis demonstrated that only one implant showed considerable skin retraction in the percutaneous area. Around the rest of the implants no downgrowth or only a very limited epidermal downgrowth was observed. Furthermore, the fiber mesh material showed good biocompatible behavior. These results confirm the previous finding that sintered titanium fiber-web structures are effective for the stabilization of percutaneous devices located in soft tissues and can improve their performance. (C) 1994 John Wiley and Sons, Inc.
引用
收藏
页码:1047 / 1054
页数:8
相关论文
共 28 条
[1]   PRESENT CLINICAL-APPLICATIONS OF OSSEOINTEGRATED PERCUTANEOUS IMPLANTS [J].
ALBREKTSSON, T ;
BRANEMARK, PI ;
JACOBSSON, M ;
TJELLSTROM, A .
PLASTIC AND RECONSTRUCTIVE SURGERY, 1987, 79 (05) :721-730
[2]   The soft tissue barrier at implants and teeth [J].
Berglundh, T. ;
Lindhe, J. ;
Ericsson, I. ;
Marinello, C. P. ;
Liljenberg, B. ;
Thomsen, P. .
CLINICAL ORAL IMPLANTS RESEARCH, 1991, 2 (02) :81-90
[3]   THE BIOPHYSICAL AND HISTOLOGIC PROPERTIES OF CAPSULES FORMED BY SMOOTH AND TEXTURED SILICONE IMPLANTS IN THE RABBIT [J].
BERN, S ;
BURD, A ;
MAY, JW .
PLASTIC AND RECONSTRUCTIVE SURGERY, 1992, 89 (06) :1037-1042
[4]  
BUSER D, 1989, Z ZAHNARZTL IMPLANTO, V5, P15, DOI DOI 10.1111/J.1600-0501.2006.01380.X
[5]   THE ROLE OF CONNECTIVE-TISSUE IN INHIBITING EPITHELIAL DOWNGROWTH ON TITANIUM-COATED PERCUTANEOUS IMPLANTS [J].
CHEHROUDI, B ;
GOULD, TRL ;
BRUNETTE, DM .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1992, 26 (04) :493-515
[6]  
DUCHEYNE P, 1984, CEMENTLESS FIXATION, P109
[7]  
DUCHEYNE P, 1983, ASTM STP, V796, P265
[8]   ELECTRON-MICROSCOPE INVESTIGATION OF SOFT-TISSUE INGROWTH INTO DACRON VELOUR WITH DOGS [J].
FELDMAN, DS ;
HULTMAN, SM ;
COLAIZZO, RS ;
VONRECUM, AF .
BIOMATERIALS, 1983, 4 (02) :105-111
[9]   NON-EPIDERMALLY INDUCED FAILURE MODES OF PERCUTANEOUS DEVICES [J].
FELDMAN, DS ;
VONRECUM, AF .
BIOMATERIALS, 1985, 6 (05) :352-356
[10]   DESIGN CRITERIA FOR PERCUTANEOUS DEVICES [J].
GROSSESIESTRUP, C ;
AFFELD, K .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1984, 18 (04) :357-382