摘要
A method of controlling the specific absorption rate (SAR) distribution in radiofrequency hyperthermia is proposed. The superposition of the current density associated with capacitively coupled electrodes and that associated with H-field coupled inductive aperture-type applicators modifies the actual current distribution in the heating material. Using a two-dimensional finite element method, we have shown that "focusing" is possible such that the SAR at the center of a phantom can be adjusted to be approximately twice that in superficial regions, even though the wavelength is considerably greater than the dimensions of the phantom.
引用
收藏
页码:644 / 647
页数:4
相关论文
共 16 条
[1]  
Franconi C., 1987, PHYSICS TECHNOLOGY H, P80
[2]  
HIRAOKA M, 1987, CANCER, V60, P121, DOI 10.1002/1097-0142(19870701)60:1<121::AID-CNCR2820600123>3.0.CO
[3]  
2-I
[4]   Physical Basis of RF Hyperthermia for Cancer Therapy (4) A Device for Non-Invasive Local Deep Heating in Hyperthermia Treatment [J].
Kato, Hirokazu ;
Ishida, Tetsuya .
Journal of Radiation Research, 1982, 23 (02) :228-233
[5]  
KATO H, 1983, J MICROWAVE POWER EE, V18, P331
[6]  
KATO H, 1985, International Journal of Hyperthermia, V1, P15, DOI 10.3109/02656738509029271
[7]  
KATO H, 1985, 1ST P ANN M JAP SOC, P174
[8]   A MODEL FOR IMPEDANCE DETERMINATIONS AND POWER DEPOSITION CHARACTERIZATION IN 3-ELECTRODE CONFIGURATIONS FOR CAPACITIVE RADIOFREQUENCY HYPERTHERMIA .A. IMPEDANCE DETERMINATIONS [J].
MORAND, A ;
BOLOMEY, JC .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1987, 34 (03) :217-222
[9]   A MODEL FOR IMPEDANCE DETERMINATIONS AND POWER DEPOSITION CHARACTERIZATION IN 3-ELECTRODE CONFIGURATIONS FOR CAPACITIVE RADIOFREQUENCY HYPERTHERMIA .B. CURRENT FLOW AND POWER DEPOSITION [J].
MORAND, A ;
BOLOMEY, JC .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1987, 34 (03) :223-232
[10]  
Nakamura N, 1987, Iyodenshi To Seitai Kogaku, V25, P114