TRANSVENOUS CARDIAC PACEMAKER, MURAL THROMBOSIS, AND PULMONARY EMBOLISM

被引:38
作者
REYNOLDS, J
ANSLINGER, D
YORE, R
PAINE, R
机构
[1] Department of Medicine, St. Luke's Hospital, St. Louis, MO
关键词
D O I
10.1016/0002-8703(69)90521-3
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
In a case of pacemaker failure, insertion of a second transvenous electrode into the right ventricle resulted in apparent immediate pulmonary embolism and simultaneous restoration of effective stimulation by the original malfunctioning electrode. The case provided an exceptional opportunity to recognize an offending perielectrode thrombus and to time precisely the sequence of events after pulmonary embolism. The effectiveness of anticoagulant therapy in preventing electrode interference by thrombosis was suggested. © 1969.
引用
收藏
页码:688 / +
页数:1
相关论文
共 6 条
[1]  
GOLDBERGER E, 1966, AM J CARDIOL, V17, P439
[2]   CATHETER PACING IN COMPLETE HEART BLOCK - TECHNIQUES AND COMPLICATIONS [J].
GORDON, AJ .
JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1965, 193 (13) :1091-&
[3]   ENDOCARDIAL PACING [J].
HARRIS, AM .
AMERICAN HEART JOURNAL, 1966, 72 (01) :135-&
[4]   TRANSVENOUS ELECTRICAL STIMULATION OF HEART-I [J].
SCHWEDEL, JB ;
ESCHER, DJW .
ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1964, 111 (A3) :972-&
[5]   TRANSVENOUS ELECTRICAL PACING OF HEART - RESULTS OF 96 INSERTIONS IN 78 PATIENTS [J].
YUCEOGLU, YZ ;
LUNGER, M ;
DRESDALE, DT .
AMERICAN HEART JOURNAL, 1966, 71 (01) :5-&
[6]  
ZUCKER R, 1963, JAMA-J AM MED ASSOC, V184, P549