PHYSICAL EVALUATION OF COMPUTED RADIOGRAPHY AS A MAMMOGRAPHIC X-RAY-IMAGING SYSTEM

被引:26
|
作者
WORKMAN, A
COWEN, AR
BRETTLE, DS
机构
[1] FAXIL, University of Leeds, Academic Unit of Medical Physics, Leeds General Infirmary
来源
BRITISH JOURNAL OF RADIOLOGY | 1994年 / 67卷 / 802期
关键词
D O I
10.1259/0007-1285-67-802-988
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The physical imaging properties of a computed radiography (CR) system operating under mammographic exposure conditions have been measured. These measurements include modulation transfer function (MTF), sensitometric response and noise power spectrum (NPS). These figures were used to derive signal-to-noise ratio (SNR) descriptors of the system performance. The same measures were derived for a mammographic film-screen system. The SNR properties of the CR system indicate superior low frequency performance over a wider dynamic range than the film-screen system. The spatial frequency dependent SNR properties of the CR system are, however, limited. The results of a psychophysical test support the results obtained from physical measurements. These results indicate that despite its limited sharpness, details of dimensions relevant to diagnosis in mammography (circular details 0.25 mm diameter to represent microcalcifications and 3 mm diameter to represent small masses) are reproduced using CR with SNRs which are at least comparable with those of a mammographic film-screen system.
引用
收藏
页码:988 / 996
页数:9
相关论文
共 50 条
  • [1] X-RAY-IMAGING - PROJECTION RADIOGRAPHY
    TANG, CM
    KUNDEL, HL
    RAVIN, CE
    HUANG, HK
    DESLATTES, RD
    BOYD, DP
    GUR, D
    ACADEMIC RADIOLOGY, 1995, 2 : S111 - S112
  • [2] SENSITIVE X-RAY-IMAGING SYSTEM
    PANIN, AM
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (01): : 620 - 621
  • [3] TFTR VERTICAL X-RAY-IMAGING SYSTEM
    JOHNSON, LC
    BITTER, M
    CHOUINARD, R
    VONGOELER, S
    HILL, KW
    LIEW, SL
    MCGUIRE, K
    PARE, V
    SAUTHOFF, NR
    YOUNG, KM
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1986, 57 (08): : 2133 - 2135
  • [4] DIGITAL X-RAY-IMAGING
    MYERS, MJ
    BRITISH JOURNAL OF RADIOLOGY, 1980, 53 (630): : 614 - 614
  • [5] DIGITAL X-RAY-IMAGING
    COWEN, AR
    MEASUREMENT SCIENCE AND TECHNOLOGY, 1991, 2 (08) : 691 - 707
  • [6] AN ALTERNATIVE TO X-RAY-IMAGING
    LYTLE, D
    PHOTONICS SPECTRA, 1992, 26 (12) : 148 - 148
  • [7] BACKSCATTER X-RAY-IMAGING
    BOSSI, RH
    FRIDDELL, KD
    NELSON, JM
    MATERIALS EVALUATION, 1988, 46 (11) : 1462 - 1467
  • [8] DIGITAL X-RAY-IMAGING
    MOORES, BM
    IEE PROCEEDINGS-A-SCIENCE MEASUREMENT AND TECHNOLOGY, 1987, 134 (02): : 115 - 125
  • [9] X-RAY-IMAGING TUBES
    SCHAGEN, P
    NDT INTERNATIONAL, 1981, 14 (01): : 9 - 14
  • [10] CONVENTIONAL X-RAY-IMAGING
    GARRETT, JA
    SMITHSON, PH
    IEE PROCEEDINGS-A-SCIENCE MEASUREMENT AND TECHNOLOGY, 1987, 134 (02): : 107 - 114