Repeatability and Reproducibility of Macular Thickness Measurements Using Fourier Domain Optical Coherence Tomography

被引:16
作者
Bruce, Alison [1 ]
Pacey, Ian E. [1 ]
Dharni, Poonam [1 ]
Scally, Andy J. [2 ]
Barrett, Brendan T. [1 ]
机构
[1] Univ Bradford, Bradford Sch Optometry & Vis Sci, Richmond Rd, Bradford BD7 1DP, W Yorkshire, England
[2] Univ Bradford, Sch Hlth, Inst Hlth Res, Bradford, W Yorkshire, England
关键词
Optical coherence tomography; retinal thickness; macula; image analysis; repeatability; reproducibility;
D O I
10.2174/1874364100903010010
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Aim: To evaluate repeatability and reproducibility of macular thickness measurements in visually normal eyes using the Topcon 3D OCT-1000. Methods: Phase 1 investigated scan repeatability, the effect of age and pupil dilation. Two groups (6 younger and 6 older participants) had one eye scanned 5 times pre and post-dilation by 1 operator. Phase 2 investigated between-operator, within and between-visit reproducibility. 10 participants had 1 un-dilated eye scanned 3 times on 2 separate visits by 2 operators. Results: Phase 1: No significant difference existed between repeat scans (p=0.75) and no significant difference was found pre- and post-dilation (p=0.54). In the younger group variation was low (95% limits +/- 3.62 mu m) and comparable across all retinal regions. The older group demonstrated greater variation (95% limits +/- 7.6 mu m). Phase 2: For a given retinal location, 95% confidence limits for within-operator, within-visit reproducibility was 5.16 mu m. This value increased to 5.56 mu m for the same operator over two visits and to 6.18 mu m for two operators over two visits. Conclusion: A high level repeatability, close to 6 mu m, of macular thickness measurement is possible using the 3D OCT-1000. Measured differences in macular thickness between successive visits that exceed 6 mu m in pre-presbyopic individuals are therefore likely to reflect actual structural change. OCT measures are more variable in older individuals and it is advisable to take a series of scans so that outliers can be more easily identified.
引用
收藏
页码:10 / 14
页数:5
相关论文
共 29 条
[1]   Analysis of reliability indices from Humphrey visual field tests in an urban glaucoma population [J].
Birt, CM ;
Shin, DH ;
Samudrala, V ;
Hughes, BA ;
Kim, C ;
Lee, D .
OPHTHALMOLOGY, 1997, 104 (07) :1126-1130
[2]  
BLAND M, 1995, INTRO MED STAT
[3]   Reproducibility of nerve fiber layer thickness measurements by use of optical coherence tomography [J].
Blumenthal, EZ ;
Williams, JM ;
Weinreb, RN ;
Girkin, CA ;
Berry, CC ;
Zangwill, LM .
OPHTHALMOLOGY, 2000, 107 (12) :2278-2282
[4]   Reproducibility of retinal nerve fiber thickness measurements using the stratus OCT in normal and glaucomatous eyes [J].
Budenz, DL ;
Chang, RT ;
Huang, XR ;
Knighton, RW ;
Tielsch, JM .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46 (07) :2440-2443
[5]   Normal macular thickness measurements in healthy eyes using stratus optical coherence tomography [J].
Chan, A ;
Duker, JS ;
Ko, TH ;
Fujimoto, JG ;
Schuman, JS .
ARCHIVES OF OPHTHALMOLOGY, 2006, 124 (02) :193-198
[6]   Clinical applications and new developments of optical coherence tomography: an evidence-based review [J].
Chen, Jennifer ;
Lee, Lawrence .
CLINICAL AND EXPERIMENTAL OPTOMETRY, 2007, 90 (05) :317-335
[7]   Retinal assessment using optical coherence tomography [J].
Costa, Rogerio A. ;
Skaf, Mirian ;
Melo, Luiz A. S., Jr. ;
Calucci, Daniela ;
Cardillo, Jose A. ;
Castro, Jarbas C. ;
Huang, David ;
Wojtkowski, Maciej .
PROGRESS IN RETINAL AND EYE RESEARCH, 2006, 25 (03) :325-353
[8]   State-of-the-art retinal optical coherence tomography [J].
Drexler, Wolfgang ;
Fujimoto, James G. .
PROGRESS IN RETINAL AND EYE RESEARCH, 2008, 27 (01) :45-88
[9]  
Gold V., 1987, INT UNION PURE APPL
[10]   OPTICAL COHERENCE TOMOGRAPHY OF THE HUMAN RETINA [J].
HEE, MR ;
IZATT, JA ;
SWANSON, EA ;
HUANG, D ;
SCHUMAN, JS ;
LIN, CP ;
PULIAFITO, CA ;
FUJIMOTO, JG .
ARCHIVES OF OPHTHALMOLOGY, 1995, 113 (03) :325-332