Stability of Programmable Shunt Valve Settings with Simultaneous Use of the Optune Transducer Array: A Case Report

被引:3
作者
Chan, Andrew K. [1 ]
Birk, Harjus S. [2 ,3 ]
Winkler, Ethan A. [1 ]
Viner, Jennifer A. [1 ]
Taylor, Jennie W. [4 ]
McDermott, Michael W. [2 ]
机构
[1] UCSF, Neurol Surg, San Francisco, CA USA
[2] Univ Calif San Francisco, Dept Neurol Surg, San Francisco, CA 94143 USA
[3] Howard Hughes Med Inst, Chevy Chase, MD 20815 USA
[4] UCSF, Neurooncol, San Francisco, CA USA
来源
CUREUS | 2016年 / 8卷 / 07期
关键词
shunt valve; hydrocephalus; optune transducer array; glioblastoma multiforme;
D O I
10.7759/cureus.675
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The Optune (R) transducer array (Novocure Ltd., Haifa, Israel) is an FDA-approved noninvasive regional therapy that aims to inhibit the growth of glioblastoma multiforme (GBM) cells via utilization of alternating electric fields. Some patients with GBM may develop hydrocephalus and benefit from subsequent shunt placement, but special attention must be paid to patients in whom programmable valves are utilized, given the potential effect of the magnetic fields on valve settings. We present the first case report illustrating the stability of programmable shunt valve settings in a neurosurgical patient undergoing therapy with the Optune device. In this study, shunt valve settings were stable over a period of five days despite Optune therapy. This is reassuring for patients with GBM who require simultaneous treatment with both the Optune device and a programmable shunt system.
引用
收藏
页数:6
相关论文
共 7 条
[1]  
[Anonymous], 2013, MEDTRONIC STRATA ADJ
[2]  
[Anonymous], 2015, NEW PHASE 3 DATA SHO
[3]   What next for newly diagnosed glioblastoma? [J].
Domingo-Musibay, Evidio ;
Galanis, Evanthia .
FUTURE ONCOLOGY, 2015, 11 (24) :3273-3283
[4]   Predicting the electric field distribution in the brain for the treatment of glioblastoma [J].
Miranda, Pedro C. ;
Mekonnen, Abeye ;
Salvador, Ricardo ;
Basser, Peter J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (15) :4137-4147
[5]   Electromagnetic field hazards involving adjustable shunt valves in hydrocephalus [J].
Schneider, T ;
Knauff, U ;
Nitsch, J ;
Firsching, R .
JOURNAL OF NEUROSURGERY, 2002, 96 (02) :331-334
[6]   Maintenance Therapy With Tumor-Treating Fields Plus Temozolomide vs Temozolomide Alone for Glioblastoma A Randomized Clinical Trial [J].
Stupp, Roger ;
Taillibert, Sophie ;
Kanner, Andrew A. ;
Kesari, Santosh ;
Steinberg, David M. ;
Toms, Steven A. ;
Taylor, Lynne P. ;
Lieberman, Frank ;
Silvani, Antonio ;
Fink, Karen L. ;
Barnett, Gene H. ;
Zhu, Jay-Jiguang ;
Henson, John W. ;
Engelhard, Herbert H. ;
Chen, Thomas C. ;
Tran, David D. ;
Sroubek, Jan ;
Iran, Nam D. ;
Hottinger, Andreas F. ;
Landolfi, Joseph ;
Desai, Rajiv ;
Caroli, Manuela ;
Kew, Yvonne ;
Honnorat, Jerome ;
Idbaih, Ahmed ;
Kirson, Eilon D. ;
Weinberg, Uri ;
Palti, Yoram ;
Hegi, Monika E. ;
Ram, Zvi .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2015, 314 (23) :2535-2543