摘要
The relatively high incidence of acetabular component failure shown in long-term followup studies of cemented total hip replacement has led to the use of noncemented fixation on the acetabular side of the total hip arthroplasty. In the first generation of noncemented acetabular devices, large-size femoral heads requiring thinner polyethylene were used with many of the popular designs, These components were also modular, but the design considerations involved in minimizing polyethylene debris were not fully appreciated, The newer designs of noncemented acetabular components are addressing the need to minimize wear and the production of particulate debris at the bearing surface and metal-polyethylene interface, Newer techniques also have been developed to provide initial component stability and bone-prosthesis contact, With these improvements, the results of noncemented acetabular fixation may outperform the results with cemented fixation. This may be especially true when considering the ease of insertion and versatility of noncemented acetabular components.
引用
收藏
页码:76 / 84
页数:9
相关论文
共 43 条
[1]  
BARRACK R, 1983, J BONE JOINT SURG BR, V75, P688
[2]  
BEAUCHESNE R, 1992, 59TH P AM AC ORTH SU, P334
[3]   DISSOCIATION OF ACETABULAR COMPONENTS AFTER TOTAL HIP-ARTHROPLASTY - REPORT OF 4 CASES [J].
BRIEN, WW ;
SALVATI, EA ;
WRIGHT, TM ;
NELSON, CL ;
HUNGERFORD, DS ;
GILLIAM, DL .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (10) :1548-1550
[4]   THE UNCEMENTED POROUS-COATED ANATOMIC TOTAL HIP-PROSTHESIS - 2-YEAR RESULTS OF A PROSPECTIVE CONSECUTIVE SERIES [J].
CALLAGHAN, JJ ;
DYSART, SH ;
SAVORY, CG .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (03) :337-346
[5]  
CALLAGHAN JJ, 1994, 61ST P AM AC ORTH SU, P249
[6]  
Cook S D, 1988, J Arthroplasty, V3, P249
[7]   THE INITIAL STABILITY OF UNCEMENTED ACETABULAR COMPONENTS [J].
CURTIS, MJ ;
JINNAH, RH ;
WILSON, VD ;
HUNGERFORD, DS .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (03) :372-376
[8]   CEMENTLESS ACETABULAR COMPONENTS [J].
ENGH, CA ;
GRIFFIN, WL ;
MARX, CL .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1990, 72 (01) :53-59
[9]   HISTOLOGICAL AND RADIOGRAPHIC ASSESSMENT OF WELL-FUNCTIONING POROUS-COATED ACETABULAR COMPONENTS - A HUMAN POSTMORTEM RETRIEVAL STUDY [J].
ENGH, CA ;
ZETTLSCHAFFER, KF ;
KUKITA, Y ;
SWEET, D ;
VIRGINIA, A ;
JASTY, M ;
BRAGDON, C .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1993, 75A (06) :814-824
[10]  
FEHRING T, 1992, 59TH P AM AC ORTH SU, P83