COMPARISON OF FLUID WARMER PERFORMANCE DURING SIMULATED CLINICAL CONDITIONS

被引:14
作者
PATEL, N [1 ]
SMITH, CE [1 ]
PINCHAK, AC [1 ]
机构
[1] CASE WESTERN RESERVE UNIV,METROHLTH MED CTR,DEPT ANESTHESIA,CLEVELAND,OH 44109
来源
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE | 1995年 / 42卷 / 07期
关键词
EQUIPMENT; FLUID WARMERS; TEMPERATURE; HYPOTHERMIA;
D O I
10.1007/BF03011886
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The study evaluated the warming ability and flow rates associated with four fluid warming devices during pressure driven infusion and during wide open gravity driven roller clamp infusion. Warmers tested were the Astotherm, Flotem Ile, Level 1 System 250 and a modified cardioplegia heat exchanger. Fluids tested were crystalloid red cells diluted with 200 ml, 0.9% saline, and undiluted red cells. The volume of fluid and outlet temperatures (point where iv tubing would be attached to the patient) were measured for each fluid and compared among warmers for each flow rate condition. For pressure driven infusion of red cells and crystalloid, the System 250, and modified heat exchanger delivered warmer fluids (33-35 degrees C) at higher flow rates (160-740 ml . min(-1)) than the Astotherm and Flotem (23-31 degrees C, 44-268 ml . min(-1), P < 0.05). For gravity driven infusion, the System 250 delivered the warmest fluids (33-36 degrees C, P < 0.05) compared with the modified heat exchanger (29-35 degrees C), Astotherm (26-32 degrees C) and Flotem (26-27 degrees C). In conclusion, the modified heat exchanger and System 250 were moderately effective (outlet temperature >32 degrees C) in warming crystalloid and red cells at pressure driven flow rates. Only the System 250 warmed red cells >35 degrees C at gravity driven flow rates. The Flotem and Astotherm were not effective in warming rapidly infused solutions. None of the warmers tested was able to deliver fluids at normothermia (>36.5 degrees C).
引用
收藏
页码:636 / 642
页数:7
相关论文
共 14 条
[1]   TEMPERATURE RELATIONSHIP TO DISTANCE AND FLOW-RATE OF WARMED-IV FLUIDS [J].
FARIES, G ;
JOHNSTON, C ;
PRUITT, KM ;
PLOUFF, RT .
ANNALS OF EMERGENCY MEDICINE, 1991, 20 (11) :1198-1200
[2]   UNINTENTIONAL HYPOTHERMIA IS ASSOCIATED WITH POSTOPERATIVE MYOCARDIAL-ISCHEMIA [J].
FRANK, SM ;
BEATTIE, C ;
CHRISTOPHERSON, R ;
NORRIS, EJ ;
PERLER, BA ;
WILLIAMS, GM ;
GOTTLIEB, SO ;
MEINERT, C ;
ROCK, P ;
PARKER, S ;
YATES, H ;
WILLIAMS, GM ;
BRESLOW, M ;
ROSENFELD, B ;
TAYLOR, D ;
BRASFIELD, B ;
BOURKE, D ;
BEZIRDJIAN, P ;
PAUL, S ;
ACHUFF, S ;
BUCHMAN, T ;
HEITMILLER, E ;
NYHAN, D ;
SITZMAN, J ;
STEVENSON, RJ .
ANESTHESIOLOGY, 1993, 78 (03) :468-476
[3]   CONTINUOUS ARTERIOVENOUS REWARMING - EXPERIMENTAL RESULTS AND THERMODYNAMIC MODEL SIMULATION OF TREATMENT FOR HYPOTHERMIA [J].
GENTILELLO, LM ;
CORTES, V ;
MOUJAES, S ;
VIAMONTE, M ;
MALININ, TL ;
HO, CH ;
GOMEZ, GA .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1990, 30 (12) :1436-1449
[4]  
GENTILELLO LM, 1992, J TRAUMA, V32, P16
[5]   BODY-TEMPERATURE AND ANESTHESIA [J].
HALL, GM .
BRITISH JOURNAL OF ANAESTHESIA, 1978, 50 (01) :39-44
[6]  
HARDY JD, 1943, J NUTR, V21, P383
[7]  
HENDRICKX HHL, 1982, ANESTH ANALG, V61, P393
[8]   HYPOTHERMIA IN TRAUMA VICTIMS - AN OMINOUS PREDICTOR OF SURVIVAL [J].
JURKOVICH, GJ ;
GREISER, WB ;
LUTERMAN, A ;
CURRERI, PW .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1987, 27 (09) :1019-1024
[9]   EVALUATION OF A BLOOD WARMER THAT UTILIZES A 40-DEGREES-C HEAT-EXCHANGER [J].
KRUSKALL, MS ;
PACINI, DG ;
MALYNN, ER ;
BUTTON, LN .
TRANSFUSION, 1990, 30 (01) :7-10
[10]   HYPOTHERMIA IN THE ELDERLY [J].
MORRISON, RC .
INTERNATIONAL ANESTHESIOLOGY CLINICS, 1988, 26 (02) :124-133