THE NEED FOR QUALITY-CONTROL IN MEASUREMENT OF MIXED VENOUS OXYGEN-SATURATION

被引:6
作者
BEARDS, SC
EDWARDS, JD
NIGHTINGALE, P
机构
[1] Intensive Care Unit, University Hospital of South Manchester, Manchester, M20 8LR, Nell Lane, West Didsbury
关键词
EQUIPMENT; CO-OXIMETERS;
D O I
10.1111/j.1365-2044.1994.tb04267.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We investigated the variability of arterial, mixed venous and peripheral venous oxygen saturation readings in three co-oximeters in regular use on two Intensive Care Units. Over a 2-week period, 96 readings were obtained from each of the following machines; two AVL 912 co-oxylite co-oximeters (Machines A and B, Biomedical Instruments, Graz, Australia) and one lL-282 (Machine C, Instrumentation Laboratories, Lexington, MA, USA). The coefficient of variation bl the arterial oxygen saturations was extremely small on all three machines (< 0.43%). The coefficient of variation in the mixed venous samples, however, ranged from 1.96 to 4.61 % on machine A, and from 2.73 to 4.71% on machine B, but only from 0.17 to 1.47% on machine C. The variation in mixed venous saturations obtained from a single blood sample repeatedly analysed on machines A and B was large enough to influence clinical management.
引用
收藏
页码:886 / 888
页数:3
相关论文
共 6 条
[1]  
KWANT G, 1989, CLIN CHEM, V35, P773
[2]  
LEDINGHAM IM, 1989, TXB CRITICAL CARE, P205
[3]   CONTINUOUS VENOUS OXIMETRY IN SURGICAL PATIENTS [J].
NELSON, LD .
ANNALS OF SURGERY, 1986, 203 (03) :329-333
[4]   BIAS AND PRECISION OF PULSE OXIMETERS AND ARTERIAL OXIMETERS [J].
NICKERSON, BG ;
SARKISIAN, C ;
TREMPER, K .
CHEST, 1988, 93 (03) :515-517
[5]  
SHOEMAKER WC, 1985, SURG CLIN N AM, V65, P811
[6]   ARTERIAL OXYGEN STATUS DETERMINED WITH ROUTINE PH/BLOOD GAS EQUIPMENT AND MULTIWAVELENGTH HEMOXIMETRY - REFERENCE VALUES, PRECISION, AND ACCURACY [J].
SIGGAARDANDERSEN, O ;
WIMBERLEY, PD ;
FOGHANDERSEN, N ;
GOTHGEN, IH .
SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1990, 50 :57-66