THE NEED FOR QUALITY-CONTROL IN MEASUREMENT OF MIXED VENOUS OXYGEN-SATURATION

被引:6
作者
BEARDS, SC
EDWARDS, JD
NIGHTINGALE, P
机构
[1] Intensive Care Unit, University Hospital of South Manchester, Manchester, M20 8LR, Nell Lane, West Didsbury
关键词
EQUIPMENT; CO-OXIMETERS;
D O I
10.1111/j.1365-2044.1994.tb04267.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We investigated the variability of arterial, mixed venous and peripheral venous oxygen saturation readings in three co-oximeters in regular use on two Intensive Care Units. Over a 2-week period, 96 readings were obtained from each of the following machines; two AVL 912 co-oxylite co-oximeters (Machines A and B, Biomedical Instruments, Graz, Australia) and one lL-282 (Machine C, Instrumentation Laboratories, Lexington, MA, USA). The coefficient of variation bl the arterial oxygen saturations was extremely small on all three machines (< 0.43%). The coefficient of variation in the mixed venous samples, however, ranged from 1.96 to 4.61 % on machine A, and from 2.73 to 4.71% on machine B, but only from 0.17 to 1.47% on machine C. The variation in mixed venous saturations obtained from a single blood sample repeatedly analysed on machines A and B was large enough to influence clinical management.