ENZYME ELECTRODE WITH COLLAGEN-IMMOBILIZED CHOLESTEROL OXIDASE FOR THE MICRO-DETERMINATION OF FREE-CHOLESTEROL

被引:35
|
作者
BERTRAND, C
COULET, PR
GAUTHERON, DC
机构
[1] Laboratoire de Biologie et Technologie des Membranes du CNRS, Universite Claude Bernard, 69622, Villeurbanne, Lyon I 43
来源
ANALYTICAL LETTERS PART B-CLINICAL AND BIOCHEMICAL ANALYSIS | 1979年 / 12卷 / 14期
关键词
Clinical Analysis; Collagen membrane; Enzymatic cholesterol; Enzyme Electrode;
D O I
10.1080/00032717908082547
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Cholesterol oxidase has been immobilized on collagen films associated to an electrochemical sensor to form a “cholesterol electrode”. The electrode poised at a potential of + 650 my vs Ag/AgCl detects the hydrogen peroxide produced in the enzymatic reaction. This device presents a very high sensitivity and a wide range of linearity (10 -7 M - 0.8.10 -4 M). The use of a non enzymatic electrode associated with the enzymatic one allowed the detection and correction of electrochemical interferences when applied to human sera for free cholesterol determination. © 1979, Taylor & Francis Group, LLC. All rights reserved.
引用
收藏
页码:1477 / 1488
页数:12
相关论文
共 1 条