A Study on Error Characteristics of Large Size Electromagnetic Flowmeter in the Range of Low Velocity

被引:3
|
作者
Lee, Dong Keun
Park, Jong Ho
机构
关键词
Electromagnetic Flowmeter; Error Characteristics; Low Velocity;
D O I
10.3795/KSME-B.2008.32.3.235
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
The large size electromagnetic flowmeter was tested to investigate the variation of its error characteristics in the range of low velocity under 0.6 m/s using flowmeter calibration system. For the two case of valve opening rate 100 % and 50 %, these tests were undertaken three times each for twelve velocity condition from 0.05 (m)/s to 0.6 (m)/s with increment of 0.05 (m)/s. It is shown that error characteristic of electromagnetic flowmeter was stabilized within +/- 0.4 % of rate both higher than 0.25 (m)/s of velocity condition and 50 % of valve opening position. But, measurement deviation of flowmeter for Phi 400 mm and Phi 600 mm was out of expected deviation range. It is necessary to correction with calibration. In conclusion, error characteristic of electromagnetic flowmeter wasn't changed proportion to its size.
引用
收藏
页码:235 / 240
页数:6
相关论文
共 50 条
  • [1] Electromagnetic flowmeter evaluation in real facilities: Velocity profiles and error analysis
    Sotero Salustiano Martim, Andre Luis
    Dalfre Filho, Jose Gilberto
    Lemos De Lucca, Yvone de Faria
    Borri Genovez, Ana Ines
    FLOW MEASUREMENT AND INSTRUMENTATION, 2019, 66 : 44 - 49
  • [2] Using an Electromagnetic Flowmeter for Ultra-low Velocity Measurement
    Fiala, P.
    Steinbauer, M.
    Szabo, Z.
    2019 PHOTONICS & ELECTROMAGNETICS RESEARCH SYMPOSIUM - SPRING (PIERS-SPRING), 2019, : 3767 - 3771
  • [3] Related technologies of narrowband filter low-velocity flow electromagnetic flowmeter
    Niu, Bin
    Chen, Song-Jing
    Pang, Bing
    Liang, Yuan-Hua
    Harbin Gongye Daxue Xuebao/Journal of Harbin Institute of Technology, 2009, 41 (05): : 148 - 151
  • [4] The Performance Characteristics of Electromagnetic Flowmeter in Vertical Low-Velocity Oil-Water Two-Phase Flow
    Jin, Ningde
    Yu, Chuang
    Han, Yunfeng
    Yang, Qiuyi
    Ren, Yingyu
    Zhai, Lusheng
    IEEE SENSORS JOURNAL, 2021, 21 (01) : 464 - 475
  • [5] Large velocity capture range and low temperatures with cavities
    Murr, K.
    PHYSICAL REVIEW LETTERS, 2006, 96 (25)
  • [6] Flow Signal Characteristics of Small Scale Electromagnetic Flowmeter in Low Conductivity Fluid Measurement
    Lim, Ki Won
    Jung, Sung Soo
    TRANSACTIONS OF THE KOREAN SOCIETY OF MECHANICAL ENGINEERS B, 2016, 40 (09) : 613 - 620
  • [7] CATHETER-TIP ELECTROMAGNETIC FLOWMETER FOR VELOCITY FLOW IN LARGE AND SMALL VESSELS AND TOTAL FLOW IN APPROPRIATE BRANCHES
    STEIN, PD
    SCHUETTE, WH
    CIRCULATION, 1968, 38 (4S6) : V189 - &
  • [8] Validation range study of a low airflow velocity thermistor sensor
    Cabrita, Admesio
    Mendes, Ricardo
    Quintela, Divo A.
    INSTRUMENTATION SCIENCE & TECHNOLOGY, 2025,
  • [9] Study on match filtering based on sequency spectrum characteristics of the Walsh transform for electromagnetic flowmeter signal measurement
    Li, Bin
    Zhang, Sheng-Nan
    Chen, Jie
    Zhao, Wang-cheng
    MEASUREMENT, 2024, 225
  • [10] A study on the vibration characteristics of a large size Korean bell
    Lee, JM
    Kim, SH
    Lee, SJ
    Jeong, JD
    Choi, HG
    JOURNAL OF SOUND AND VIBRATION, 2002, 257 (04) : 779 - 790