EIT IMAGE-RECONSTRUCTION USING SENSITIVITY WEIGHTED FILTERED BACKPROJECTION

被引:44
作者
KOTRE, CJ
机构
[1] Regional Med. Phys. Dept., Newcastle Gen. Hospital
关键词
D O I
10.1088/0967-3334/15/2A/017
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A set of specifications thought useful to advance the utility of electrical impedance tomography as a clinical imaging modality is proposed, together with an approach to image reconstruction designed to fulfil these specifications. Computer simulations are used to investigate the performance of this reconstruction approach on simple, known, impedance distributions and to develop the method to the stage of a complete reconstruction algorithm. The algorithm is then tested on data sets produced by measurements on a physical phantom, and on a set of measurements made on a volunteer human subject
引用
收藏
页码:A125 / A136
页数:12
相关论文
共 7 条
[1]  
Barber DC., 1986, INFORMATION PROCESSI, P106
[2]  
Brown B H, 1987, Clin Phys Physiol Meas, V8 Suppl A, P91, DOI 10.1088/0143-0815/8/4A/012
[4]   A FAST APPROXIMATION FOR THE CALCULATION OF POTENTIAL DISTRIBUTIONS IN ELECTRICAL-IMPEDANCE TOMOGRAPHY [J].
KOTRE, CJ .
CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT, 1988, 9 (04) :353-361
[5]  
KOTRE CJ, 1993, THESIS U NEWCASTLE U
[6]  
Seagar A D, 1987, Clin Phys Physiol Meas, V8 Suppl A, P13, DOI 10.1088/0143-0815/8/4A/003
[7]  
WEBB S, 1988, PHYSICS MED IMAGING