The isobaric effect correction and measurement of the Ru isotopes by thermal ionization mass spectrometry

被引:0
|
作者
Jeon, Young-Shin [1 ]
Kim, Jung-Suk [1 ]
Han, Sun-Ho [1 ]
Song, Kyuseok [1 ]
机构
[1] Korea Atom Energy Res Inst, 150,Duckjin Dong, Daejeon, South Korea
来源
ANALYTICAL SCIENCE AND TECHNOLOGY | 2010年 / 23卷 / 05期
关键词
ruthenium; isotope measurement; thermal ionization mass spectrometry; isobaric effect correction;
D O I
10.5806/AST.2010.23.5.498
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
It is very difficult to get stable ion peak intensity of ruthenium by thermal ionization mass spectrometry because of its high ionization potential and high volatility of its oxides which causes to lose much of ruthenium ions, so the intensity of the signal decrease quickly. Accordingly, a study was performed in oder to increase the ionization efficiency and to prevent sample losses due to volatilization and to check with isobaric effect by impurities in filament for the measurement of ruthenium isotopes. Both single filament and double filament were tested. The former was proved to be more efficient for the stable and strong intensity of signal and revealed less isobaric effect from the molybdenum (Mo) as a filament impurity. Also, when the temperature of filament increased too high, the isobaric effect from Mo greatly appeared. That is, Mo impurity from filament gave a serious effect for measuring the ruthenium isotopes. It was proved to be of importance that filament current should be slowly increased with time interval. Finally, ruthenium isotopes were accurately measured by correction with measuring Mo-94/(99) Ru.
引用
收藏
页码:498 / 504
页数:7
相关论文
共 50 条
  • [1] The isobaric effect on the measurement of Gd isotopes by thermal ionization mass spectrometry
    Jeon, Young-Shin
    Park, Jong-Ho
    Joe, Kih-Soo
    Han, Sun-Ho
    Song, Kyuseok
    ANALYTICAL SCIENCE AND TECHNOLOGY, 2009, 22 (05): : 415 - 421
  • [2] Precise measurement of zirconium isotopes by thermal ionization mass spectrometry
    Sahoo, SK
    Masuda, A
    CHEMICAL GEOLOGY, 1997, 141 (1-2) : 117 - 126
  • [3] Precise measurement of chlorine isotopes by thermal ionization mass spectrometry
    Neskovic, OM
    Veljkovic, NV
    Velickovic, SR
    Deric, AJ
    Miljevic, NR
    Golobocanin, DD
    NUKLEONIKA, 2002, 47 : S85 - S87
  • [4] Measurement of boron isotopes by negative thermal ionization mass spectrometry
    LIU Wei-Guo1
    WANG Zhao-Rong(Department of Earth and Space Sciences
    Key Laboratory of Loess and Quaternary Geology
    Nuclear Science and Techniques, 2000, (02) : 79 - 84
  • [5] Accurate measurement of ruthenium isotopes by negative thermal ionization mass spectrometry
    Huang, M
    Liu, YZ
    Masuda, A
    ANALYTICAL CHEMISTRY, 1996, 68 (05) : 841 - 844
  • [6] Thermal ionization mass spectrometry of molybdenum isotopes
    Wieser, ME
    De Laeter, JR
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2000, 197 : 253 - 261
  • [7] Uranium Isobaric Effect Correction for Determination of Pu-238 in Spent Nuclear Fuel by Thermal Ionization Mass Spectrometry
    Jeon, Young-Shin
    Park, Yang-Soon
    Kim, Jung-Suk
    Park, Yong Joon
    Song, Kyuseok
    ASIAN JOURNAL OF CHEMISTRY, 2014, 26 (13) : 4052 - 4054
  • [8] Application of Total Evaporation Technique Using Thermal Ionization Mass Spectrometry in Measurement of Plutonium Isotopes
    Ying Z.-C.
    Xiong C.-J.
    Zhang L.-H.
    Li H.-B.
    Hao X.-J.
    Journal of Chinese Mass Spectrometry Society, 2021, 42 (03) : 305 - 313
  • [9] Study in the Measurement of Boron Isotopes by Thermal Ionization Mass Spectrometer
    Jeon, Young Shin
    Joe, Kih Soo
    Han, Sun Ho
    Park, Yong Joon
    Jee, Kwang Yong
    ANALYTICAL SCIENCE AND TECHNOLOGY, 2005, 18 (01): : 27 - 34
  • [10] SIMULTANEOUS MEASUREMENT OF LITHIUM AND BORON ISOTOPES AS LITHIUM TETRABORATE ION BY THERMAL IONIZATION MASS-SPECTROMETRY
    SAHOO, SK
    MASUDA, A
    ANALYST, 1995, 120 (02) : 335 - 339