DIRECT MICROSEQUENCE ANALYSIS OF POLYPEPTIDES USING AN IMPROVED SEQUENATOR, A NON-PROTEIN CARRIER (POLYBRENE), AND HIGH-PRESSURE LIQUID-CHROMATOGRAPHY

被引:371
作者
HUNKAPILLER, MW [1 ]
HOOD, LE [1 ]
机构
[1] CALTECH,DIV BIOL,PASADENA,CA 91125
关键词
D O I
10.1021/bi00604a016
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
We have combined the use of a nonprotein carrier (Polybrene), high pressure liquid chromatography, and modifications in Edman chemistry with the improvements of a commercial spinning cup sequenator suggested by Witt-mann-Liebold [Wittmann-Liebold, B. (1973) Hoppe-Seyler's Z. Physiol. Chem. 354, 1415] to analyze amino acid phenyl-thiohydantoins obtained from automated Edman degradation of microquantities of polypeptide directly without the use of radiolabel. This approach has allowed us to determine the sequence of the N-terminal 47 residues of sperm whale myo-globin starting with 200 pmol of protein, 77 residues of an antibody light chain with 5 nmol of protein, and 54 residues of an antibody heavy chain with 8 nmol of protein. In addition, we completely sequenced a hydrophobic 14-residue peptide at the 1.5-nmol level. Our technique of direct analysis for microsamples is capable of providing routine, extended N-terminal sequence analysis for nanomole and subnanomole levels of polypeptides and proteins, and it also is applicable to analysis of more classical sample quantities. © 1978, American Chemical Society. All rights reserved.
引用
收藏
页码:2124 / 2133
页数:10
相关论文
共 26 条
[1]   HLA MEMBRANE ANTIGENS - SEQUENCING BY INTRINSIC RADIOACTIVITY [J].
BALLOU, B ;
MCKEAN, DJ ;
FREEDLENDER, EF ;
SMITHIES, O .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1976, 73 (12) :4487-4491
[2]   APPLICATION OF 0.1 M-QUADROL TO MICROSEQUENCE OF PROTEINS AND SEQUENCE OF TRYPTIC PEPTIDES [J].
BRAUER, AW ;
MARGOLIES, MN ;
HABER, E .
BIOCHEMISTRY, 1975, 14 (13) :3029-3035
[3]   ISOLATION AND N-TERMINAL AMINO-ACID SEQUENCE OF MEMBRANE-BOUND HUMAN HLA-A AND HLA-B ANTIGENS [J].
BRIDGEN, J ;
SNARY, D ;
CRUMPTON, MJ ;
BARNSTABLE, C ;
GOODFELLOW, P ;
BODMER, WF .
NATURE, 1976, 261 (5557) :200-205
[4]   SOLID-PHASE SEQUENCING - NEW SUPPORT FOR HIGH SENSITIVITY DEGRADATION OF PEPTIDES AND PROTEINS [J].
BRIDGEN, J .
FEBS LETTERS, 1975, 50 (02) :159-162
[5]   AMINO-ACID SEQUENCE OF NH2-TERMINAL EXTRA PIECE SEGMENTS OF PRECURSORS OF MOUSE IMMUNOGLOBULIN LAMBDA-1-TYPE AND KAPPA-TYPE LIGHT-CHAINS [J].
BURSTEIN, Y ;
SCHECHTER, I .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1977, 74 (02) :716-720
[6]   AUTOMATIC PROCEDURES FOR DETERMINING AMINO-ACID SEQUENCES OF PEPTIDES [J].
CREWTHER, WG ;
INGLIS, AS .
ANALYTICAL BIOCHEMISTRY, 1975, 68 (02) :572-585
[7]   ON THE MECHANISM OF THE PHENYL ISOTHIOCYANATE DEGRADATION OF PEPTIDES [J].
EDMAN, P .
ACTA CHEMICA SCANDINAVICA, 1956, 10 (05) :761-768
[8]   A PROTEIN SEQUENATOR [J].
EDMAN, P ;
BEGG, G .
EUROPEAN JOURNAL OF BIOCHEMISTRY, 1967, 1 (01) :80-&
[9]   APPLICATION OF SEQUENATOR ANALYSES TO STUDY OF PROTEINS [J].
HERMODSON, MA ;
NEURATH, H ;
WALSH, KA ;
TITANI, K ;
ERICSSON, LH .
BIOCHEMISTRY, 1972, 11 (24) :4493-+
[10]  
JACOBS JW, 1975, J BIOL CHEM, V250, P3629