MEASUREMENT OF VANADIUM, NICKEL, AND ARSENIC IN SEAWATER AND URINE REFERENCE MATERIALS BY INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY WITH CRYOGENIC DESOLVATION

被引:52
作者
ALVES, LC
ALLEN, LA
HOUK, RS
机构
[1] IOWA STATE UNIV SCI & TECHNOL,US DOE,AMES LAB,AMES,IA 50011
[2] IOWA STATE UNIV SCI & TECHNOL,DEPT CHEM,AMES,IA 50011
关键词
D O I
10.1021/ac00066a011
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Addition of a small dose (2%) of H-2 to the aerosol gas flow enhanced analyte signals by a factor of 2-3, which compensated for the loss of analyte signal that accompanied earlier efforts at cryogenic desolvation with inductively coupled plasma mass spectrometry (ICP-MS). Vanadium, nickel, and arsenic at microgram per liter levels in urine, river, and seawater reference materials were determined. The polyatomic ions ClO+, CaO+, and ArCl+, which normally cause severe overlap interferences for these elements, were attenuated to manageable levels by cryogenic desolvation. The samples were simply diluted with 1% HNO3 so that the chloride could be removed as HCl. The analytical results obtained for these standard reference materials agreed closely with the certified or recommended values. The detection limit ranges (3sigma) obtained were 10-1000 ng L-1 for V, 0.03-20 mug L-1 for Ni, and 4-7000 ng L-1 for As in the original samples. The samples were introduced by flow injection to minimize clogging of the sampling orifice.
引用
收藏
页码:2468 / 2471
页数:4
相关论文
共 30 条
[1]   REDUCTION OF POLYATOMIC ION INTERFERENCES IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY BY CRYOGENIC DESOLVATION [J].
ALVES, LC ;
WIEDERIN, DR ;
HOUK, RS .
ANALYTICAL CHEMISTRY, 1992, 64 (10) :1164-1169
[2]   DETERMINATION OF TRACE-METALS IN REFERENCE WATER STANDARDS BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY WITH ONLINE PRECONCENTRATION [J].
BEAUCHEMIN, D ;
BERMAN, SS .
ANALYTICAL CHEMISTRY, 1989, 61 (17) :1857-1862
[4]   INVESTIGATIONS ON MIXED-GAS PLASMAS PRODUCED BY ADDING NITROGEN TO THE PLASMA GAS IN ICP-MS [J].
BEAUCHEMIN, D ;
CRAIG, JM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1991, 46 (05) :603-614
[5]   INDUCTIVELY COUPLED PLASMA AS AN ION-SOURCE FOR HIGH-RESOLUTION MASS-SPECTROMETRY [J].
BRADSHAW, N ;
HALL, EFH ;
SANDERSON, NE .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1989, 4 (08) :801-803
[6]   DETERMINATION OF ARSENIC BY HYDRIDE GENERATION INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY USING A TUBULAR MEMBRANE GAS-LIQUID SEPARATOR [J].
BRANCH, S ;
CORNS, WT ;
EBDON, L ;
HILL, S ;
ONEILL, P .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1991, 6 (02) :155-158
[7]   EFFECT OF ORGANIC-SOLVENTS AND MOLECULAR GASES ON POLYATOMIC ION INTERFERENCES IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
EVANS, EH ;
EBDON, L .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1990, 5 (06) :425-429
[8]  
EVANS EH, 1984, ANAL AT SPECTROM, V4, P299
[9]   MASS-SPECTRA AND IONIZATION TEMPERATURES IN AN ARGON-NITROGEN INDUCTIVELY COUPLED PLASMA [J].
HOUK, RS ;
MONTASER, A ;
FASSEL, VA .
APPLIED SPECTROSCOPY, 1983, 37 (05) :425-428
[10]   TRACE-METAL ISOTOPIC ANALYSIS OF MICROLITER SOLUTION VOLUMES BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
HOUK, RS ;
THOMPSON, JJ .
BIOMEDICAL MASS SPECTROMETRY, 1983, 10 (02) :107-112