Dosimetric, Radiobiological and Secondary Cancer Risk Evaluation in Head-and-Neck Three-dimensional Conformal Radiation Therapy, Intensity-Modulated Radiation Therapy, and Volumetric Modulated Arc Therapy: A Phantom Study

被引:5
作者
Rehman, Jalil Ur [1 ,3 ,5 ]
Isa, Muhammad [4 ,6 ,7 ]
Ahmad, Nisar [1 ]
Nasar, Gulfam [2 ]
Asghar, H. M. Noor Ul Huda Khan [1 ]
Gilani, Zaheer Abbas [1 ]
Chow, James C. L. [6 ,7 ]
Afzal, Muhammad [3 ]
Ibbott, Geoffrey S. [5 ]
机构
[1] Balochistan Univ Informat Technol Engn & Manageme, Dept Phys, Quetta, Pakistan
[2] Balochistan Univ Informat Technol Engn & Manageme, Dept Chem, Quetta, Pakistan
[3] Islamia Univ Bahawalpur, Dept Phys, Bahawalpur, Pakistan
[4] Univ Gujrat, Dept Phys, Hafiz Hayat Campus, Gujrat, Pakistan
[5] Univ Texas MD Anderson Canc Ctr, Houston, TX 77030 USA
[6] Univ Toronto, Dept Radiat Oncol, Toronto, ON, Canada
[7] Princess Margaret Canc Ctr, Radiat Med Program, Toronto, ON, Canada
关键词
EBT2; FILM; imaging and radiation oncology core; intensity-modulated radiation therapy; three-dimensional conformal radiation therapy; thermoluminescent dosimeter; volumetric modulated arc therapy;
D O I
10.4103/jmp.JMP_106_17
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This analysis estimated secondary cancer risks after volumetric modulated arc therapy (VMAT) and compared those risks to the risks associated with other modalities of head-and-neck (H&N) radiotherapy. Images of H&N anthropomorphic phantom were acquired with a computed tomography scanner and exported via digital imaging and communications in medicine (DICOM) standards to a treatment planning system. Treatment plans were performed using a VMAT dual-arc technique, a nine-field intensity-modulated radiation therapy (IMRT) technique, and a four-field three-dimensional conformal therapy (3DCRT) technique. The prescription dose was 66.0 Gy for all three techniques, but to accommodate the range of dosimeter responses, we delivered a single dose of 6.60 Gy to the isocenter. The lifetime risk for secondary cancers was estimated according to National Council on Radiation Protection and Measurements (NCRP) Report 116. VMAT delivered the lowest maximum doses to esophagus (23 Gy), and normal brain (40 Gy). In comparison, maximum doses for 3DCRT were 74% and 40%, higher than those for VMAT for the esophagus, and normal brain, respectively. The normal tissue complication probability and equivalent uniform dose for the brain (2.1%, 0.9%, 0.8% and 3.8 Gy, 2.6 Gy, 2.3 Gy) and esophagus (4.2%, 0.7%, 0.4% and 3.7 Gy, 2.2 Gy, 1.8 Gy) were calculated for the 3DCRT, IMRT and VMAT respectively. Fractional esophagus OAR volumes receiving more than 20 Gy were 3.6% for VMAT, 23.6% for IMRT, and 100% for 3DCRT. The calculations for mean doses, NTCP, EUD and OAR volumes suggest that the risk of secondary cancer induction after VMAT is lower than after IMRT and 3DCRT.
引用
收藏
页码:129 / 135
页数:7
相关论文
共 50 条
[31]   Dosimetric comparison of three dimensional conformal radiation therapy versus intensity modulated radiation therapy in accelerated partial breast irradiation [J].
Moorthy, S. ;
Elhateer, H. S. ;
Majumdar, S. K. D. ;
Mohammed, S. ;
Patnaik, R. ;
Narayanamurty .
INDIAN JOURNAL OF CANCER, 2016, 53 (01) :147-151
[32]   Dental radiation dosimetric maps from intensity-modulated radiation therapy planning for head and neck cancers [J].
Polce, Simran ;
Gogineni, Emile ;
Antone, Jeffrey ;
Ghaly, Maged ;
Keith Frank, Douglas ;
Segal, Joshua D. ;
Parashar, Bhupesh .
HEAD AND NECK-JOURNAL FOR THE SCIENCES AND SPECIALTIES OF THE HEAD AND NECK, 2021, 43 (05) :1428-1439
[33]   Patterns of Care and Outcomes Associated With Intensity-Modulated Radiation Therapy Versus Conventional Radiation Therapy for Older Patients With Head-and-Neck Cancer [J].
Yu, James B. ;
Soulos, Pamela R. ;
Sharma, Richa ;
Makarov, Danil V. ;
Decker, Roy H. ;
Smith, Benjamin D. ;
Desai, Rani A. ;
Cramer, Laura D. ;
Gross, Cary P. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 83 (01) :E101-E107
[34]   Two-dimensional and Three-dimensional Image-guided Evaluation of Patient Positioning for Intensity-modulated Radiation Therapy of Head and Neck Cancer [J].
Johari, Zul Iskandar ;
Azmi, Nor Aniza ;
Ab Muin, Nur Fa'izah ;
Latiff, Rukiah A. ;
Fadzil, Muhammad Safwan Ahmad ;
Razali, Rosmizan Ahmad .
MALAYSIAN JOURNAL OF FUNDAMENTAL AND APPLIED SCIENCES, 2023, 19 (05) :915-926
[35]   Dosimetric analysis of three-dimensional conformal radiotherapy, intensity-modulated radiotherapy-step and shoot, helical tomotherapy, and volumetric modulated arc therapy in prostate cancer radiotherapy [J].
Gozal, Fenny ;
Gondhowiardjo, Soehartati Argadikoesoema ;
Kodrat, Henry ;
Wibowo, Wahyu Edy .
JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2021, 17 (04) :893-900
[36]   Dosimetric Comparison of Hypofractionated Regimen in Breast Cancer Using Two Different Techniques: Intensity-Modulated Radiation Therapy (IMRT) and Volumetric-Modulated Arc Therapy (VMAT) [J].
Prasun, Pallav ;
Kharade, Vipin ;
Pal, Vikas ;
Gupta, Manish ;
Das, Saikat ;
Pasricha, Rajesh .
CUREUS JOURNAL OF MEDICAL SCIENCE, 2023, 15 (04)
[37]   Accuracy of Real-time Couch Tracking During 3-dimensional Conformal Radiation Therapy, Intensity Modulated Radiation Therapy, and Volumetric Modulated Arc Therapy for Prostate Cancer [J].
Wilbert, Juergen ;
Baier, Kurt ;
Hermann, Christian ;
Flentje, Michael ;
Guckenberger, Matthias .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 85 (01) :237-242
[38]   Intensity modulated radiation therapy for head and neck cancer: The standard [J].
Maingon, P. ;
Crehange, G. ;
Chamois, J. ;
Khoury, C. ;
Truc, G. .
CANCER RADIOTHERAPIE, 2011, 15 (6-7) :473-476
[39]   Intensity-modulated radiation therapy in head and neck cancers: The Mallinckrodt experience [J].
Chao, KSC ;
Low, DA ;
Perez, CA ;
Purdy, JA .
INTERNATIONAL JOURNAL OF CANCER, 2000, 90 (02) :92-103
[40]   Late Toxicities of Intensity-Modulated Radiation Therapy for Head and Neck Rhabdomyosarcoma [J].
Lockney, Natalie A. ;
Friedman, Danielle Novetsky ;
Wexler, Leonard H. ;
Sklar, Charles A. ;
Casey, Dana L. ;
Wolden, Suzanne L. .
PEDIATRIC BLOOD & CANCER, 2016, 63 (09) :1608-1614