THE SPATIAL-RESOLUTION PERFORMANCE OF A TIME-RESOLVED OPTICAL IMAGING-SYSTEM USING TEMPORAL EXTRAPOLATION

被引:41
作者
HEBDEN, JC
HALL, DJ
DELPY, DT
机构
[1] Department of Medical Physics, University College London, London, WC1E 6JA
基金
英国惠康基金;
关键词
OPTICAL IMAGING; TIME-OF-FLIGHT; SPATIAL RESOLUTION; BREAST IMAGING;
D O I
10.1118/1.597457
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Optical imaging methods are being explored as a potential means of screening for breast cancer. Previous investigations of time-resolved imaging techniques have suggested that due to the lack of photons with sufficiently small pathlengths, the spatial resolution achievable through a human breast would be unlikely to be better than a centimeter. Experimental results presented here indicate, however, that higher resolution may be achieved by extrapolating the measured temporal distribution of transmitted photons. This is performed using a least-squares fit between data and an analytic model of photon transport. The spatial resolution of a time-resolved imaging system was evaluated by measuring the edge response produced by an opaque mask embedded in the center of a 51-mm-thick, very highly scattering medium. The limiting spatial resolution was improved from about 13 mm to about 5 mm. © 1995, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:201 / 208
页数:8
相关论文
共 24 条
[1]   TIME-RESOLVED TRANSILLUMINATION FOR MEDICAL DIAGNOSTICS [J].
ANDERSSONENGELS, S ;
BERG, R ;
SVANBERG, S ;
JARLMAN, O .
OPTICS LETTERS, 1990, 15 (21) :1179-1181
[3]  
CHANCE B, 1993, P SOC PHOTO OPT INST, V1888
[4]  
Cutler M., 1929, SURG GYNECOL OBSTET, V48, P721
[5]   TIME-GATED IMAGING THROUGH SCATTERING MEDIA USING STIMULATED RAMAN AMPLIFICATION [J].
DUNCAN, MD ;
MAHON, R ;
TANKERSLEY, LL ;
REINTJES, J .
OPTICS LETTERS, 1991, 16 (23) :1868-1870
[6]  
FORREST P, 1986, BREAST CANCER SCREEN
[7]   PHOTON PATH-LENGTH DISTRIBUTIONS FOR TRANSMISSION THROUGH OPTICALLY TURBID SLABS [J].
GANDJBAKHCHE, AH ;
WEISS, GH ;
BONNER, RF ;
NOSSAL, R .
PHYSICAL REVIEW E, 1993, 48 (02) :810-818
[8]   RESOLUTION LIMITS FOR OPTICAL TRANSILLUMINATION OF ABNORMALITIES DEEPLY EMBEDDED IN TISSUES [J].
GANDJBAKHCHE, AH ;
NOSSAL, R ;
BONNER, RF .
MEDICAL PHYSICS, 1994, 21 (02) :185-191
[9]   EVALUATING THE SPATIAL-RESOLUTION PERFORMANCE OF A TIME-RESOLVED OPTICAL IMAGING-SYSTEM [J].
HEBDEN, JC .
MEDICAL PHYSICS, 1992, 19 (04) :1081-1087
[10]   TIME RESOLVED IMAGING THROUGH A HIGHLY SCATTERING MEDIUM [J].
HEBDEN, JC ;
KRUGER, RA ;
WONG, KS .
APPLIED OPTICS, 1991, 30 (07) :788-794