AUTOMATED VOLTAMMETRIC SYSTEM FOR SHIPBOARD DETERMINATION OF METAL SPECIATION IN SEA-WATER

被引:61
作者
ACHTERBERG, EP [1 ]
VAN DEN BERG, CMG [1 ]
机构
[1] UNIV LIVERPOOL, OCEANOG LABS, LIVERPOOL L69 3BX, ENGLAND
关键词
VOLTAMMETRY; CHROMIUM; COPPER; METAL SPECIATION; SEA WATER; SHIPBOARD DETERMINATION; WATERS;
D O I
10.1016/0003-2670(94)85053-4
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An automated and semi-intelligent voltammetric system is described for trace metal analysis. The system consists of a voltammeter interfaced with a personal computer, a sample changer, 2 peristaltic pumps, a motor burette and a hanging mercury drop electrode. The system carries out fully automatically approximately 5 metal determinations per hour (including at least 3 repetitive scans and calibration by standard addition) at trace levels encountered in clean sea water. The computer program decides what level of standard addition to use and evaluates the data prior to switching to the next sample. Alternatively, the system can be used to carry out complexing ligand titrations with copper whilst recording the labile copper concentration; in this mode up to 8 full titrations are carried out per day. Depth profiles for chromium speciation in the Mediterranean Sea and a profile for copper complexing ligand concentrations in the North Atlantic Ocean measured on board-ship with the system are presented. The chromium speciation was determined using a new method to differentiate between Cr(In) and Cr(VI) utilizing adsorption of Cr(III) on silica particles.
引用
收藏
页码:463 / 471
页数:9
相关论文
共 50 条