WORKING RESOLUTION OF 0.010 CM-1 BETWEEN 20 CM-1 AND 1200 CM-1 BY A FOURIER SPECTROMETER

被引:56
作者
KAUPPINEN, J
机构
[1] University of Oulu, Physics Department, Oulu 10
来源
APPLIED OPTICS | 1979年 / 18卷 / 11期
关键词
D O I
10.1364/AO.18.001788
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The original double-beam Michelson-type Fourier spectrometer at the University of Oulu has been partly modified so that a maximum resolution of slightly better than 0.010 cm−1 between 20 cm−1 and 1200 cm−1 can be obtained. The resolution is determined as a halfwidth of an observed line. The accuracy of the line position is 0.0005 cm−1 in favorable conditions. The Fourier spectrometer is connected on-line to a minicomputer, in order to make the fast Fourier transforms of interferograms of up to 106 sampling points. The maximum optical path difference of the interferometer is about 1.5 m, giving a maximum theoretical resolution of 0.004 cm−1. © 1979 Optical Society of America.
引用
收藏
页码:1788 / 1796
页数:9
相关论文
共 37 条