NORMAL LEFT-VENTRICULAR DIMENSIONS AND FUNCTION - INTERSTUDY REPRODUCIBILITY OF MEASUREMENTS WITH CINE MR IMAGING

被引:345
|
作者
SEMELKA, RC [1 ]
TOMEI, E [1 ]
WAGNER, S [1 ]
MAYO, J [1 ]
KONDO, C [1 ]
SUZUKI, J [1 ]
CAPUTO, GR [1 ]
HIGGINS, CB [1 ]
机构
[1] UNIV CALIF SAN FRANCISCO,DEPT RADIOL,BOX 0628,C309,SAN FRANCISCO,CA 94143
关键词
Heart; ejection fraction; flow dynamics; function; MR studies; 524.1214; ventricles; volume; Magnetic resonance (MR); cine study;
D O I
10.1148/radiology.174.3.2305059
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The authors evaluated the reproducibility of measurements of ventricular dimensions obtained with cine magnetic resonance (MR) imaging performed on two occasions in 11 healthy subjects. Two reviewers analyzed the studies in a blinded fashion to determine interobserver and interstudy variability of measurements of left ventricular (LV) mass, volume, ejection fraction, and systolic wall stress. LV mass showed good reproducibility between studies, with 3.6% and 3.8% variability for LV end-systolic mass for the two observers. LV end-diastolic volume varied by 5.2% and 3.9%, and LV end-systolic volume, by 9.7% and 9.0%. Variability for LV ejection fraction was 5.0% and 4.9%. The largest interstudy variability was end-systolic wall stress, 11.1% and 13.5%, which was due mostly to changes in arterial pressure between the two studies. It is concluded that anatomic and functional measurements from cine MR images are reproducible between studies. The small interstudy variability is likely related to the fact that the measurements are derived directly from cine MR images that encompass the entire heart rather than depend on measurements of only sample images and the use of geometric models.
引用
收藏
页码:763 / 768
页数:6
相关论文
共 50 条
  • [31] Improving the reproducibility of MR-derived left ventricular volume and function measurements with a semi-automatic threshold-based segmentation algorithm
    Jaspers, Karolien
    Freling, Hendrik G.
    van Wijk, Kees
    Romijn, Elisabeth I.
    Greuter, Marcel J. W.
    Willems, Tineke P.
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2013, 29 (03) : 617 - 623
  • [32] Quantitative assessment of left ventricular function with interactive real-time spiral and radial MR imaging
    Spuentrup, E
    Schroeder, J
    Mahnken, AH
    Schaeffter, T
    Botnar, RM
    Kühl, HP
    Hanrath, P
    Günther, RW
    Buecker, A
    RADIOLOGY, 2003, 227 (03) : 870 - 876
  • [33] Comparison of interstudy reproducibility of equilibrium electrocardiography-gated SPECT radionuclide angiography versus planar radionuclide angiography for the quantification of global left ventricular function
    Daou D.
    Coaguila C.
    Benada A.
    Journal of Nuclear Cardiology, 2006, 13 (2) : 233 - 243
  • [34] MEASUREMENT OF REGIONAL LEFT-VENTRICULAR FUNCTION USING LABELED MAGNETIC-RESONANCE-IMAGING
    BUCHALTER, MB
    SIMS, C
    DIXON, AK
    WEISSBERG, PL
    STONE, DL
    SHAH, NJ
    HALL, LD
    WILKINSON, ID
    LOMAS, DJ
    FREER, CEL
    BRITISH JOURNAL OF RADIOLOGY, 1991, 64 (766) : 953 - 958
  • [35] ECHOCARDIOGRAPHIC AND HEMODYNAMIC INDEXES OF LEFT-VENTRICULAR PRELOAD IN PATIENTS WITH NORMAL AND ABNORMAL VENTRICULAR-FUNCTION
    CHEUNG, AT
    SAVINO, JS
    WEISS, SJ
    AUKBURG, SJ
    BERLIN, JA
    ANESTHESIOLOGY, 1994, 81 (02) : 376 - 387
  • [36] INTRAMURAL MYOCARDIAL SHORTENING IN HYPERTENSIVE LEFT-VENTRICULAR HYPERTROPHY WITH NORMAL PUMP FUNCTION
    PALMON, LC
    REICHEK, N
    YEON, SB
    CLARK, NR
    BROWNSON, D
    HOFFMAN, E
    AXEL, L
    CIRCULATION, 1994, 89 (01) : 122 - 131
  • [37] NORMAL AND DIFFUSELY ABNORMAL MYOCARDIUM IN HUMANS - FUNCTIONAL AND METABOLIC CHARACTERIZATION WITH P-31 MR SPECTROSCOPY AND CINE MR IMAGING
    AUFFERMANN, W
    CHEW, WM
    WOLFE, CL
    TAVARES, NJ
    PARMLEY, WW
    SEMELKA, RC
    DONNELLY, T
    CHATTERJEE, K
    HIGGINS, CB
    RADIOLOGY, 1991, 179 (01) : 253 - 259
  • [38] CALCULATION OF 3-DIMENSIONAL LEFT-VENTRICULAR STRAINS FROM BIPLANAR TAGGED MR IMAGES
    MOORE, CC
    ODELL, WG
    MCVEIGH, ER
    ZERHOUNI, EA
    JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1992, 2 (02): : 165 - 175
  • [39] Comparison of left and right ventricular ejection and filling parameters by fast cine MR imaging in breath-hold technique: Clinical study of 42 patients with cardiomyopathy and coronary heart disease.
    Rominger, MB
    Bachmann, GF
    Geuer, M
    Puzik, M
    Ricken, WW
    Rau, WS
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1999, 170 (06): : 534 - 541
  • [40] Assessment of left ventricular function by breath-hold cine MR imaging: Comparison of different steady-state free precession sequences
    Kunz, RP
    Oellig, F
    Krummenauer, F
    Oberholzer, K
    Romaneehsen, B
    Vomweg, TW
    Horstick, G
    Hayes, C
    Thelen, M
    Kreitner, KF
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2005, 21 (02) : 140 - 148