AUTOMATED MULTIPLE FLOW INJECTION ANALYSIS FOR FLAME ATOMIC-ABSORPTION SPECTROMETRY

被引:112
作者
WOLF, WR
STEWART, KK
机构
[1] Nutrient Composition Laboratory, Nutrition Institute, Maryland 20705, USDA, SEA, HNC, Beitsvilie
关键词
D O I
10.1021/ac50044a024
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Nonsegmented continuous flow analysis, termed Flow Injection Analysis (FIA), has been adapted for the automation of sample introduction in flame atomic absorption spectroscopy. Automated Multiple Flow Injection Analysis (AMFIA) entails insertion of discrete micro volumes of liquid sample into a constantly flowing solvent stream which is pumped directly into the nebulizer of an atomic absorption spectrometer. This allows analysis of discrete micro volume samples (25 to 300 μL), with high precision (2% RSD), low detection limits (3 ng for Zn and 4 ng for Cu) and speed of analysis (120 to 180 samples per hour). Identical results are obtained on zinc and copper content of food digests and NBS Standard Reference Materials with the AMFIA system and analysis by direct aspiration of the same sample solutions. © 1979, American Chemical Society. All rights reserved.
引用
收藏
页码:1201 / 1205
页数:5
相关论文
共 18 条
  • [1] BETTERIDGE D, 1978, ANAL CHEM, V50, pA832, DOI 10.1021/ac50031a001
  • [2] FISHER CG, 1976, AT ABSORPT NEWSL, V17, P33
  • [3] MICROSAMPLING NEBULIZER TECHNIQUE FOR PREMIXED FLAME ATOMIC SPECTROMETRY
    FRY, RC
    NORTHWAY, SJ
    DENTON, MB
    [J]. ANALYTICAL CHEMISTRY, 1978, 50 (12) : 1719 - 1722
  • [4] JACKWORTH E, 1975, SPECTROCHIM ACTA B, V30, P169
  • [5] Manning D. C., 1975, Atomic Absorption Newsletter, V14, P99
  • [6] MENIS O, 1970, ANAL FLAME SPECTROSC
  • [7] ROCKLAND LB, J FOOD SCI
  • [8] RUZICKA J, 1977, Patent No. 4022575
  • [9] RUZICKA J, 1974, Patent No. 484684
  • [10] RUZICKA J, 1975, ANAL CHIM ACTA, V78, P17