Artificial urinary sphincters for male stress urinary incontinence: current perspectives

被引:35
作者
Cordon, Billy H. [1 ]
Singla, Nirmish [1 ]
Singla, Ajay K. [2 ]
机构
[1] Univ Texas Southwestern Med Ctr Dallas, Dept Urol, Dallas, TX 75390 USA
[2] Univ Toledo, Coll Med, Dept Urol, 3000 Arlington Ave,MS 1091, Toledo, OH 43614 USA
关键词
artificial urinary sphincter; stress urinary incontinence; post prostatectomy incontinence; prostheses and implants; review;
D O I
10.2147/MDER.S93637
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The artificial urinary sphincter (AUS), which has evolved over many years, has become a safe and reliable treatment for stress urinary incontinence and is currently the gold standard. After 4 decades of existence, there is substantial experience with the AUS. Today AUS is most commonly placed for postprostatectomy stress urinary incontinence. Only a small proportion of urologists routinely place AUS. In a survey in 2005, only 4% of urologists were considered high-volume AUS implanters, performing >20 per year. Globally, similar to 11,500 AUSs are placed annually. Over 400 articles have been published regarding the outcomes of AUS, with a wide variance in success rates ranging from 61% to 100%. Generally speaking, the AUS has good long-term outcomes, with social continence rates of similar to 79% and high patient satisfaction usually between 80% and 90%. Despite good outcomes, a substantial proportion of patients, generally similar to 25%, will require revision surgery, with the rate of revision increasing with time. Complications requiring revision include infection, urethral atrophy, erosion, and mechanical failure. Most infections are gram-positive skin flora. Urethral atrophy and erosion lie on a spectrum resulting from the same problem, constant urethral compression. However, these two complications are managed differently. Mechanical failure is usually a late complication occurring on average later than infection, atrophy, or erosions. Various techniques may be used during revisions, including cuff relocation, downsizing, transcorporal cuff placement, or tandem cuff placement. Patient satisfaction does not appear to be affected by the need for revision as long as continence is restored. Additionally, AUS following prior sling surgery has comparable outcomes to primary AUS placement. Several new inventions are on the horizon, although none have been approved for use in the US at this point.
引用
收藏
页码:175 / 183
页数:9
相关论文
共 79 条
[11]   Surgery for postprostatectomy incontinence: which procedure for which patient? [J].
Comiter, Craig .
NATURE REVIEWS UROLOGY, 2015, 12 (02) :91-99
[12]   Surgery Insight: surgical management of postprostatectomy incontinence - the artificial urinary sphincter and male sling [J].
Comiter, Craig V. .
NATURE CLINICAL PRACTICE UROLOGY, 2007, 4 (11) :615-624
[13]   The artificial urinary sphincter and male sling for postprostatectomy incontinence: Which patient should get which procedure? [J].
Comiter, Craig V. ;
Dobberfuhl, Amy D. .
INVESTIGATIVE AND CLINICAL UROLOGY, 2016, 57 (01) :3-13
[14]   A PRELIMINARY-REPORT ON A NEW HYDRAULIC SPHINCTER FOR CONTROLLING URINARY-INCONTINENCE [J].
CRAGGS, MD ;
CHAFFEY, NJ ;
MUNDY, AR .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1991, 15 (02) :58-62
[15]  
Crivellaro S, 2015, NEUROUROL URODYN
[16]   The Impact of an Antibiotic Coating on the Artificial Urinary Sphincter Infection Rate [J].
de Cogain, Mitra R. ;
Elliott, Daniel S. .
JOURNAL OF UROLOGY, 2013, 190 (01) :113-117
[17]   Tandem cuff artificial urinary sphincter as a salvage procedure following failed primary sphincter placement for the treatment of post-prostatectomy incontinence [J].
DiMarco, DS ;
Elliott, DS .
JOURNAL OF UROLOGY, 2003, 170 (04) :1252-1254
[18]   Mayo Clinic long-term analysis of the functional durability of the AMS 800 artificial urinary sphincter: A review of 323 cases [J].
Elliott, DS ;
Barrett, DM .
JOURNAL OF UROLOGY, 1998, 159 (04) :1206-1208
[19]   Revision Techniques After Artificial Urinary Sphincter Failure in Men: Results From a Multicenter Study [J].
Eswara, Jairam R. ;
Chan, Robert ;
Vetter, Joel M. ;
Lai, H. Henry ;
Boone, Timothy B. ;
Brandes, Steven B. .
UROLOGY, 2015, 86 (01) :176-180
[20]   Systematic Review and Meta-analysis of Studies Reporting Urinary Continence Recovery After Robot-assisted Radical Prostatectomy [J].
Ficarra, Vincenzo ;
Novara, Giacomo ;
Rosen, Raymond C. ;
Artibani, Walter ;
Carroll, Peter R. ;
Costello, Anthony ;
Menon, Mani ;
Montorsi, Francesco ;
Patel, Vipul R. ;
Stolzenburg, Jens-Uwe ;
Van der Poel, Henk ;
Wilson, Timothy G. ;
Zattoni, Filiberto ;
Mottrie, Alexandre .
EUROPEAN UROLOGY, 2012, 62 (03) :405-417