NAFION-COATED CARBON-FIBER ELECTRODES FOR NEUROCHEMICAL STUDIES IN BRAIN-TISSUE

被引:174
作者
CAPELLA, P [1 ]
GHASEMZADEH, B [1 ]
MITCHELL, K [1 ]
ADAMS, RN [1 ]
机构
[1] UNIV KANSAS,DEPT CHEM,LAWRENCE,KS 66045
关键词
D O I
10.1002/elan.1140020303
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Extensive studies of the characteristics of electrochemically pretreated, Nafion‐coated carbon fibers used for in vivo electrochemistry are summarized. While electrochemical pretreatment provides high sensitivity for the monitoring of catecholamine neurotransmitters, it causes fibers to respond slowly. The ionomer film, which confers selectivity, adds relatively little to the electrode time constant. In terms of sensitivity and time constant changes, serious interactions occur during the course of the in vivo experiments while the electrodes are exposed to brain tissue. Fortunately, selectivity characteristics do not change during in vivo experiments. Postcalibration procedures and other measurement techniques that provide reliable experimental measurements are discussed. Copyright © 1990 VCH Publishers, Inc.
引用
收藏
页码:175 / 182
页数:8
相关论文
共 19 条