QUANTITATION OF GLUCOSE-CONCENTRATION USING A GLUCOSE OXIDASE CATALASE ELECTRODE BY POTENTIOMETRIC MEASUREMENT

被引:29
|
作者
LIU, CC
WINGARD, LB
WOLFSON, SK
YAO, SJ
DRASH, AL
SCHILLER, JG
机构
[1] UNIV PITTSBURGH,SCH MED,DEPT PHARMACOL,PITTSBURGH,PA 15260
[2] UNIV PITTSBURGH,SCH MED,DEPT NEUROL SURG,PITTSBURGH,PA 15260
[3] UNIV PITTSBURGH,SCH MED,DEPT PEDIAT,PITTSBURGH,PA 15260
来源
BIOELECTROCHEMISTRY AND BIOENERGETICS | 1979年 / 6卷 / 01期
基金
美国国家卫生研究院;
关键词
D O I
10.1016/0302-4598(79)85003-5
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
Glucose oxidase and catalase were immobilized in polyacrylamide gel around a platinum screen and used to measure the concentration of glucose in 0.1 M phosphate buffer, pH 7.3, at 37°C. This enzyme electrode produced a direct potentiometric signal with reference to a standard Ag|AgCl electrode. The potential difference was linearly related to the logarithm of the concentration of glucose over the range of 3-40 cg/dm3 (mg/dl). The electrode was stable for at least 32 hours at 37°C. Hydrogen peroxide, generated by the oxidation of glucose and depleted by decomposition and diffusion, was the apparent source of the potentiometric signal. Direct potentiometric measurement has the advantages of simple readout, continuous detection, and small electrode size, and does not require external polarising potential. This approach should be suitable for the continuous in vivo monitoring of blood glucose levels if the upper limit of specificity for glucose (presently 40 mg/dl) can be extended. © 1979.