Preliminary performance characterization of DbPET2.1, a PET scanner dedicated to the imaging of the breast and extremities

被引:10
作者
Ferrero, Andrea [1 ]
Peng, Qiyu [2 ]
Burkett, George W., Jr. [3 ]
Sumanasena, Buddika [3 ]
Moses, William W. [2 ]
Badawi, Ramsey D. [1 ,3 ]
机构
[1] Univ Calif Davis, Biomed Engn Grad Grp, Davis, CA 95616 USA
[2] Lawrence Berkeley Natl Lab, Lawrence, KS USA
[3] UCDavis Med Ctr, Dept Radiol, Sacramento, CA USA
来源
BIOMEDICAL PHYSICS & ENGINEERING EXPRESS | 2015年 / 1卷 / 01期
关键词
breast PET/CT; high resolution PET; breast cancer;
D O I
10.1088/2057-1976/1/1/015202
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The combined effort of several laboratories at our institution resulted in the building of the first high-resolution PET/CT prototype dedicated to imaging the extremities of the body. Ongoing clinical trials for breast-cancer diagnosis and assessment of the response to treatment underlined the need for a second-generation prototype with improved electronics and spatial resolution. Apreliminary version has been assembled and fully characterized. In this work, we present further improvements in detector performance, as well as the readout electronics for the PET component. The detector consists of a 16 x 16 array of 1.27 x 1.27 x 20 mm(3) lutetium yttrium orthosilicate crystals, the smallest crystal size for completed breast PET prototypes to date, directly coupled to a position-sensitive photomultiplier tube (PSPMT). The scintillator crystals are polished on all six faces and separated by a similar to 70 mu m enhanced specular reflector. The readout electronics were redesigned to reduce their footprint and improve timing resolution. Wereport a detector energy and timing resolution of 12% and 1.0 ns, respectively, and an average intrinsic spatial resolution of 1.29mm(central row in one detector array). The new PET/CT has been fully assembled, and initial system characterization is being performed. We report a system energy resolution of 15.7%, a timing resolution of 1.5 ns and a filtered back-projection image spatial resolution in the center of the field of view of 1.6 mm.
引用
收藏
页数:9
相关论文
共 22 条
  • [1] Design and evaluation of the Clear-PEM scanner for positron emission mammography
    Abreu, MC
    Aguiar, JD
    Almeida, FG
    Almeida, P
    Bento, P
    Carrico, B
    Ferreira, M
    Ferreira, NC
    Gonçalves, F
    Leong, C
    Lopes, F
    Lousa, P
    Martins, MV
    Matela, N
    Mendes, PR
    Moura, R
    Nobre, J
    Oliveira, N
    Ortigao, C
    Peralta, L
    Pereira, R
    Rego, J
    Ribeiro, R
    Rodrigues, P
    Sampaio, J
    Santos, AI
    Silva, L
    Silva, JC
    Sousa, P
    Teixeira, IC
    Teixeira, JP
    Trindade, A
    Varela, J
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (01) : 71 - 77
  • [2] Agarwal S, 2013, WORLD MOL IM C OR PR
  • [3] Characterization of the Clear-PEM Breast Imaging Scanner Performance
    Albuquerque, E.
    Almeida, F. G.
    Almeida, P.
    Auffray, E.
    Barbosa, J.
    Bastos, A. L.
    Bexiga, V.
    Bugalho, R.
    Cardoso, C.
    Carmona, S.
    Carneiro, J. F.
    Carrico, B.
    Ferreira, C. S.
    Ferreira, N. C.
    Ferreira, M.
    Frade, M.
    Goncalves, F.
    Guerreiro, C.
    Lecoq, P.
    Leong, C.
    Lousa, P.
    Machado, P.
    Martins, M. V.
    Matela, N.
    Moura, R.
    Neves, J.
    Neves, P.
    Oliveira, N.
    Ortigao, C.
    Piedade, F.
    Pinheiro, J. F.
    Relvas, P.
    Rivetti, A.
    Rodrigues, P.
    Rolo, I.
    Santos, A. I.
    Santos, J.
    Silva, M. M.
    Tavernier, S.
    Teixeira, I. C.
    Teixeira, J. P.
    Silva, J. C.
    Silva, R.
    Trindade, A.
    Varela, J.
    [J]. 2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 3487 - +
  • [4] [Anonymous], 2008, 42008 NU NEMA
  • [5] An X-ray Computed Tomography/Positron Emission Tomography System Designed Specifically for Breast Imaging
    Boone, John M.
    Yang, Kai
    Burkett, George W.
    Packard, Nathan J.
    Huang, Shih-ying
    Bowen, Spencer
    Badawi, Ramsey D.
    Lindfors, Karen K.
    [J]. TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2010, 9 (01) : 29 - 43
  • [6] Initial Characterization of a Dedicated Breast PET/CT Scanner During Human Imaging
    Bowen, Spencer L.
    Wu, Yibao
    Chaudhari, Abhijit J.
    Fu, Lin
    Packard, Nathan J.
    Burkett, George W.
    Yang, Kai
    Lindfors, Karen K.
    Shelton, David K.
    Hagge, Rosalie
    Borowsky, Alexander D.
    Martinez, Steve R.
    Qi, Jinyi
    Boone, John M.
    Cherry, Simon R.
    Badawi, Ramsey D.
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2009, 50 (09) : 1401 - 1408
  • [7] Chaudhari AJ, 2011, ARTHRITIS RHEUM-US, V63, pS76
  • [8] Ferrero A, 2011, J NUCL MED, V52, P661
  • [9] Ferrero A, 2014, SOC NUCL MED, V55, P2144
  • [10] Ferrero A, 2011, IEEE NUCL SCI CONF R, P3256, DOI 10.1109/NSSMIC.2011.6152585