COMPARISON OF NONSPECIFIC RADIOIMMUNOASSAY, HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY, AND FLUORESCENCE POLARIZATION IMMUNOASSAY FOR CYCLOSPORINE MONITORING IN RENAL-TRANSPLANTATION

被引:8
|
作者
LEE, SF
YANG, WC
SHANN, TY
LUI, WY
WANG, RB
机构
[1] VET GEN HOSP TAIPEI,DIV NEPHROL,TAIPEI 11217,TAIWAN
[2] VET GEN HOSP TAIPEI,DEPT GEN SURG,TAIPEI 11217,TAIWAN
关键词
CYCLOSPORINE; RADIOIMMUNOASSAY; HIGH-PERFORMANCE LIQUID CHROMATOGRAPHY; FLUORESCENCE POLARIZATION IMMUNOASSAY;
D O I
10.1097/00007691-199103000-00011
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Three methods, i.e., nonspecific radioimmunoassay (RIA; Incstar), fluorescence polarization immunoassay (FPIA; TDx Abbott), and high-performance liquid chromatography (HPLC), have been used for monitoring cyclosporine blood levels in renal transplantation patients. The levels obtained from 135 samples showed a modest correlation between RIA and HPLC, FPIA and HPLC, RIA and FRIA. The mean ratios of RIA to HPLC, FPIA to HPLC, and RIA to FPIA were 2.96, 4.14, and 0.73. The significant variations in cyclosporine levels result from the cross-reaction of antibody with some cyclosporine metabolites, by which these two methods often overestimate the true blood cyclosporine level. HPLC is a more effective and reliable method for pharmacokinetic studies and blood level monitoring of cyclosporine in clinical practice.
引用
收藏
页码:152 / 156
页数:5
相关论文
共 50 条
  • [11] COMPARISON OF CYCLOSPORINE MEASUREMENT IN WHOLE-BLOOD BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY, MONOCLONAL FLUORESCENCE POLARIZATION IMMUNOASSAY, AND MONOCLONAL ENZYME MULTIPLIED IMMUNOASSAY
    DUSCI, LJ
    HACKETT, LP
    CHISWELL, GM
    ILETT, KF
    THERAPEUTIC DRUG MONITORING, 1992, 14 (04) : 327 - 332
  • [12] Tacrolimus monitoring in renal transplantation: A comparison between high-performance liquid chromatography and immunoassay
    Borrows, R
    Chusney, G
    Loucaidou, M
    Singh, S
    James, A
    Stichbury, J
    Van Tromp, J
    Cairns, T
    Griffith, M
    Hakim, N
    McLean, A
    Palmer, A
    Papalois, V
    Taube, D
    TRANSPLANTATION PROCEEDINGS, 2005, 37 (04) : 1733 - 1735
  • [13] THERAPEUTIC CYCLOSPORINE MONITORING - COMPARISON OF RADIOIMMUNOASSAY AND HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY METHODS IN ORGAN TRANSPLANT RECIPIENTS
    HIRVISALO, EL
    KIVISTO, KT
    NEUVONEN, PJ
    THERAPEUTIC DRUG MONITORING, 1990, 12 (04) : 353 - 358
  • [14] COMPARISON OF HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH FLUORESCENCE POLARIZATION IMMUNOASSAY FOR THE ANALYSIS OF VANCOMYCIN IN PATIENTS WITH CHRONIC-RENAL-FAILURE
    NAJJAR, TA
    ALDHUWAILIE, AA
    TEKLE, A
    JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1995, 672 (02): : 295 - 299
  • [15] PHARMACOKINETICS OF CYCLOSPORINE ESTIMATED BY RADIOIMMUNOASSAY AND HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY
    PLEBANI, M
    MASIERO, M
    DELFAVERO, E
    BONVICINI, P
    BURLINA, A
    CLINICAL CHEMISTRY, 1988, 34 (06) : 1271 - 1271
  • [16] SPECIFIC MONOCLONAL RADIOIMMUNOASSAY AND FLUORESCENCE POLARIZATION IMMUNOASSAY FOR TROUGH CONCENTRATION AND AREA-UNDER-THE-CURVE MONITORING OF CYCLOSPORINE IN RENAL-TRANSPLANTATION
    LINDHOLM, A
    NAPOLI, K
    RUTZKY, L
    KAHAN, BD
    THERAPEUTIC DRUG MONITORING, 1992, 14 (04) : 292 - 300
  • [17] CYCLOSPORINE MONITORING IN KIDNEY-TRANSPLANT PATIENTS - COMPARISON BETWEEN RADIOIMMUNOASSAY AND HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY METHODS
    MOJAVERIAN, P
    GREEN, PJ
    CARABALLO, H
    FERGUSON, RK
    DRUG INTELLIGENCE & CLINICAL PHARMACY, 1985, 19 (06): : 454 - 454
  • [18] DIBEKACIN ASSAY IN SERUM BY AUTOMATED FLUORESCENCE POLARIZATION IMMUNOASSAY (ABBOTT TDX) - COMPARISON WITH HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY, SUBSTRATE-LABELED FLUORESCENT IMMUNOASSAY AND RADIOIMMUNOASSAY
    ROLLMAN, B
    VANDERAUWERA, P
    TULKENS, PM
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1986, 4 (01) : 53 - 61
  • [19] A comparison of high-performance liquid chromatography and fluorescence polarization immunoassay for therapeutic drug monitoring of tricyclic antidepressants
    Hackett, LP
    Dusci, LJ
    Ilett, KF
    THERAPEUTIC DRUG MONITORING, 1998, 20 (01) : 30 - 34
  • [20] CYCLOSPORINE MONITORING IN KIDNEY-TRANSPLANTED CHILDREN - HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY VERSUS RADIOIMMUNOASSAY
    KHOSS, AE
    HAMILTON, G
    ILLIEVICH, V
    MUHLBACHER, F
    ROTH, E
    STEGER, H
    WLADIKA, W
    HOWANIETZ, H
    BALZAR, E
    CHILD NEPHROLOGY AND UROLOGY, 1989, 9 (1-2) : 42 - 45