BIOMECHANICAL ANALYSIS OF BONE-MINERAL DENSITY, INSERTION TECHNIQUE, SCREW TORQUE, AND HOLDING STRENGTH OF ANTERIOR CERVICAL PLATE SCREWS

被引:68
|
作者
RYKEN, TC
CLAUSEN, JD
TRAYNELIS, VC
GOEL, VK
机构
[1] UNIV IOWA,DIV NEUROSURG,IOWA CITY,IA
[2] UNIV IOWA,DEPT BIOMED ENGN,IOWA CITY,IA 52242
关键词
BIOMECHANICS; BONE MINERAL DENSITY; CERVICAL SPINE; TORQUE INSTRUMENTATION;
D O I
10.3171/jns.1995.83.2.0324
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The bone mineral density (BMD) of 99 cadaveric cervical vertebral bodies (C3-7) was determined using dual x-ray absorptiometry. The vertebral bodies were randomly assigned to receive either a unicortical (51 bodies) or bicortical (48 bodies) Caspar cervical plating screw. The initial insertion torque was measured using a digital electronic torque wrench, and the force required to withdraw the screw from the vertebral body was determined. The mean BMD for the total group of 99 was 0.787 +/- 0.154 g/cm(2), the mean insertion torque was 0.367 +/- 0.243 newton-meters, and the mean pullout force was 210.4 +/- 158.1 newtons. A significant correlation was noted between BMD and torque (p < 0.0001, r = 0.42), BMD and pullout force (p < 0.0001, r = 0.54), and torque and pullout force (p < 0.0001, r = 0.88). Although the BMD of the unicortical and bicortical groups was equivalent (p = 0.92), the insertion torque and pullout force differed significantly (p = 0.02 and p = 0.008, respectively) for the unicortical and bicortical groups. A holding index for each screw and insertion technique was defined as the product of the BMD and insertion torque. The calculated holding index and resultant pullout force were significantly correlated for both techniques of screw insertion (r = 0.92), and a significant difference in holding index was observed with unicortical versus bicortical screw placement (p = 0.04). The determination of BMD and measurement of insertion torque to create a unique holding index provides an assessment of bone-screw interaction and holding strength of the screw, both of which impact on the resultant stability of cervical instrumentation. As the number of cervical plating systems increases, the determination of a holding index for various screws and insertion techniques may assist in the comparison of cervical instrumentation.
引用
收藏
页码:324 / 329
页数:6
相关论文
共 18 条
  • [1] Biomechanical evaluation of relationship of screw pullout strength, insertional torque, and bone mineral density in the cervical spine
    Reitman, CA
    Nguyen, L
    Fogel, GR
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2004, 17 (04): : 306 - 311
  • [2] STRENGTH OF ANTERIOR VERTEBRAL SCREW FIXATION IN RELATIONSHIP TO BONE-MINERAL DENSITY
    LIM, TH
    AN, HS
    EVANICH, C
    HASANOGLU, KY
    MCGRADY, L
    WILSON, CR
    JOURNAL OF SPINAL DISORDERS, 1995, 8 (02): : 121 - 125
  • [3] Anterior cervical plate fixation: A biomechanical study to evaluate the effects of plate design, endplate preparation, and bone mineral density
    Dvorak, MF
    Pitzen, T
    Zhu, QG
    Gordon, JD
    Fisher, CG
    Oxland, TR
    SPINE, 2005, 30 (03) : 294 - 301
  • [4] Does cancellous screw insertion torque depend on bone mineral density and/or microarchitecture?
    Ab-Lazid, Rosidah
    Perilli, Egon
    Ryan, Melissa K.
    Costi, John J.
    Reynolds, Karen J.
    JOURNAL OF BIOMECHANICS, 2014, 47 (02) : 347 - 353
  • [5] Insufficient stability of pedicle screws in osteoporotic vertebrae: biomechanical correlation of bone mineral density and pedicle screw fixation strength
    Lukas Weiser
    Gerd Huber
    Kay Sellenschloh
    Lennart Viezens
    Klaus Püschel
    Michael M. Morlock
    Wolfgang Lehmann
    European Spine Journal, 2017, 26 : 2891 - 2897
  • [6] ASSESSMENT OF UNICORTICAL AND BICORTICAL FIXATION IN A QUASI-STATIC CADAVERIC MODEL - ROLE OF BONE-MINERAL DENSITY AND SCREW TORQUE
    RYKEN, TC
    GOEL, VK
    CLAUSEN, JD
    TRAYNELIS, VC
    SPINE, 1995, 20 (17) : 1861 - 1867
  • [7] Insertion torque and pullout force of rescue screws for anterior cervical plate fixation in a fatigued initial pilot hole
    Pitzen, T
    Franta, F
    Barbier, D
    Steudel, WI
    JOURNAL OF NEUROSURGERY-SPINE, 2004, 1 (02) : 198 - 201
  • [8] Insufficient stability of pedicle screws in osteoporotic vertebrae: biomechanical correlation of bone mineral density and pedicle screw fixation strength
    Weiser, Lukas
    Huber, Gerd
    Sellenschloh, Kay
    Viezens, Lennart
    Pueschel, Klaus
    Morlock, Michael M.
    Lehmann, Wolfgang
    EUROPEAN SPINE JOURNAL, 2017, 26 (11) : 2891 - 2897
  • [9] The biomechanical effect of artificial and human bone density on stopping and stripping torque during screw insertion
    Tsuji, Matthew
    Crookshank, Meghan
    Olsen, Michael
    Schemitsch, Emil H.
    Zdero, Racl
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2013, 22 : 146 - 156
  • [10] Biomechanical analysis of anterior cervical spine plate fixation systems with unicortical and bicortical screw purchase
    Wolfgang Lehmann
    Michael Blauth
    Daniel Briem
    Ulf Schmidt
    European Spine Journal, 2004, 13 : 69 - 75