A SIMPLE DATA LOSS MODEL FOR POSITRON CAMERA SYSTEMS

被引:24
作者
ERIKSSON, L
WIENHARD, K
DAHLBOM, M
机构
[1] MAX PLANCK INST NEUROL RES, COLOGNE, GERMANY
[2] UNIV CALIF LOS ANGELES, SCH MED, LOS ANGELES, CA 90024 USA
关键词
D O I
10.1109/23.322950
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A simple model to describe data losses in PET cameras is presented. The model is not intended to be used primarily for dead time corrections in existing scanners, although this is possible. Instead the model is intended to be used for data simulations in order to determine the figures of merit of future camera systems, based on data handling state-of-art solutions. The model assumes the data loss to be factorized into two components, one describing the detector or block-detector performance and the other the remaining data handling such as coincidence determinations, data transfer and data storage. Two modern positron camera systems have been investigated in terms of this model. These are the Siemens-CTI ECAT EXACT and ECAT EXACT HR systems, which both have an axial field-of-view (FOV) of about 15 cm. The both have retractable septa and can acquire data from the whole volume within the FOV and can reconstruct volume image data. An example is given how to use the model for live time calculations in a futuristic large axial FOV cylindrical system.
引用
收藏
页码:1566 / 1570
页数:5
相关论文
共 9 条
[1]  
CASEY ME, 1992, THESIS U TENNESSEE K
[2]   IMPROVED DETECTION OF FOCAL CEREBRAL BLOOD-FLOW CHANGES USING 3-DIMENSIONAL POSITRON EMISSION TOMOGRAPHY [J].
CHERRY, SR ;
WOODS, RP ;
HOFFMAN, EJ ;
MAZZIOTTA, JC .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1993, 13 (04) :630-638
[3]   UNIFIED DEADTIME CORRECTION MODEL FOR PET [J].
DAUBEWITHERSPOON, ME ;
CARSON, RE .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1991, 10 (03) :267-275
[4]   A STUDY OF DATA LOSS AND MISPOSITIONING DUE TO PILEUP IN 2-D DETECTORS IN PET [J].
GERMANO, G ;
HOFFMAN, EJ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (02) :671-675
[5]   EVALUATION OF THE PERFORMANCE-CHARACTERISTICS OF THE PC 4600 POSITRON EMISSION TOMOGRAPH [J].
KEARFOTT, KJ ;
CARROLL, LR .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1984, 8 (03) :502-513
[6]   DEAD TIME CORRECTION AND COUNTING STATISTICS FOR POSITRON TOMOGRAPHY [J].
MAZOYER, BM ;
ROOS, MS ;
HUESMAN, RH .
PHYSICS IN MEDICINE AND BIOLOGY, 1985, 30 (05) :385-399
[7]   THE EFFECT OF LIVE TIME IN COMPONENTS OF A POSITRON TOMOGRAPH ON IMAGE QUANTIFICATION [J].
THOMPSON, CJ ;
MEYER, E .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1987, 34 (01) :337-343
[8]  
WIENHARD K, IN PRESS ECAT EXACT
[9]  
YAMAMOTO S, 1986, J NUCL MED, V27, P1925