DIGITAL CHEST RADIOGRAPHY WITH PHOTOSTIMULABLE STORAGE PHOSPHORS - SIGNAL-TO-NOISE RATIO AS A FUNCTION OF KILOVOLTAGE WITH MATCHED EXPOSURE RISK

被引:60
作者
CHOTAS, HG [1 ]
FLOYD, CE [1 ]
DOBBINS, JT [1 ]
RAVIN, CE [1 ]
机构
[1] DUKE UNIV,DEPT BIOMED ENGN,DURHAM,NC 27706
关键词
LUNG; RADIOGRAPHY; BEDSIDE; DIGITAL; STORAGE PHOSPHOR; THORAX;
D O I
10.1148/radiology.186.2.8421741
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A photostimulable storage phosphor (PSP) digital radiography system was evaluated regarding the signal-to-noise ratio (S/N) on chest images acquired at differing peak kilovoltage settings but with matched risk from radiation exposure. Images of two chest phantoms were acquired by using bedside (portable) imaging equipment at tube voltages ranging from 60 to 120 kV. Exposure factors were set so that the effective dose equivalent, a risk estimator weighted for various organs, was approximately equal in all exposures. The S/N in the lung-equivalent regions was found to be slightly higher (maximum, 15%) in the low-energy exposures, while the S/N values in the mediastinum- and subdiaphragm-equivalent regions were approximately equal at all kilovoltage settings. The absence of a high sensitivity of S/N to kilovoltage in risk-matched PSP images should enable institutions to select x-ray beam quality on the basis of other imaging criteria.
引用
收藏
页码:395 / 398
页数:4
相关论文
共 15 条
  • [1] BLUME H, 1988, 1987 P CHEST IM C MA, P194
  • [2] SMALL OBJECT CONTRAST IN AMBER AND CONVENTIONAL CHEST RADIOGRAPHY
    CHOTAS, HG
    VANMETTER, RL
    JOHNSON, GA
    RAVIN, CE
    [J]. RADIOLOGY, 1991, 180 (03) : 853 - 859
  • [3] THRESHOLD PERCEPTION PERFORMANCE WITH COMPUTED AND SCREEN-FILM RADIOGRAPHY - IMPLICATIONS FOR CHEST RADIOGRAPHY
    DOBBINS, JT
    RICE, JJ
    BEAM, CA
    RAVIN, CE
    [J]. RADIOLOGY, 1992, 183 (01) : 179 - 187
  • [4] EFFECTIVE DOSE EQUIVALENTS, HE, IN DIAGNOSTIC-RADIOLOGY
    HUDA, W
    BISSESSUR, K
    [J]. MEDICAL PHYSICS, 1990, 17 (06) : 998 - 1003
  • [5] ICRP, 1977, ICRP PUBL, V26, P1383
  • [6] 2-YEAR CLINICAL-EXPERIENCE WITH A COMPUTED RADIOGRAPHY SYSTEM
    KANGARLOO, H
    BOECHAT, MI
    BARBARIC, Z
    TAIRA, RK
    CHO, PS
    MANKOVICH, NJ
    HO, BKT
    ELDREDGE, SL
    HUANG, HK
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1988, 151 (03) : 605 - 608
  • [7] PRELIMINARY EXPERIENCE WITH PORTABLE DIGITAL IMAGING FOR INTENSIVE-CARE RADIOGRAPHY
    MARGLIN, SI
    ROWBERG, AH
    GODWIN, JD
    [J]. JOURNAL OF THORACIC IMAGING, 1990, 5 (01) : 49 - 54
  • [8] Merritt C R, 1985, J Thorac Imaging, V1, P1, DOI 10.1097/00005382-198512000-00003
  • [9] MERRITT CRB, 1985, RADIOGRAPHICS, V5, P397
  • [10] MOTZ J W, 1978, Medical Physics (Woodbury), V5, P8, DOI 10.1118/1.594405