STABILIZATION OF AN INDUCTIVELY COUPLED PLASMA FOR INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY WITH A FLARED TORCH EXTENSION

被引:18
作者
GRAY, AL
机构
[1] NERC ICP-MS Facility, Department of Geology, Royal Holloway and Bedford New College, Egham, Surrey, TW20 0EX, Egham Hill
关键词
INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY; PLASMA STABILITY; ACOUSTIC NOISE; AUDIO-FREQUENCY PEAK; TORCH EXTENSION;
D O I
10.1039/ja9920701151
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An audio-frequency peak in noise power spectra of signals from emission and mass spectrometers, which use an inductively coupled plasma (ICP) as a source, has been associated by a number of workers with instability at the boundary of the plasma plume where it enters the surrounding atmosphere at the mouth of the torch. It has been shown in emission spectrometry that the extension of the outer envelope of the torch beyond the region of the plume normally viewed can eliminate this peak. Because of the need to place the ion extraction aperture just downstream of the torch mouth this does not appear to be straightforward in the case of ICP mass spectrometry but this paper reports the use of a flared torch extension that provides the necessary separation of the plasma and the surrounding atmosphere. Colour photographic evidence is provided of a major improvement of plasma stability.
引用
收藏
页码:1151 / &
相关论文
共 13 条
[1]   NOISE-POWER SPECTRA OF OPTICAL AND ACOUSTIC-EMISSION SIGNALS FROM AN INDUCTIVELY COUPLED PLASMA [J].
BELCHAMBER, RM ;
HORLICK, G .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1982, 37 (01) :17-27
[2]   NOISE POWER SPECTRAL CHARACTERISTICS OF AN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETER [J].
CRAIN, JS ;
HOUK, RS ;
ECKELS, DE .
ANALYTICAL CHEMISTRY, 1989, 61 (06) :606-612
[3]   PLASMA SOURCE-MASS SPECTROMETRY USING AN INDUCTIVELY COUPLED PLASMA AND A HIGH-RESOLUTION QUADRUPOLE MASS FILTER [J].
DATE, AR ;
GRAY, AL .
ANALYST, 1981, 106 (1269) :1255-1267
[4]   STUDIES OF A LOW-NOISE LAMINAR-FLOW TORCH FOR INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY .1. FUNDAMENTAL CHARACTERISTICS [J].
DAVIES, J ;
SNOOK, RD .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1986, 1 (03) :195-201
[5]   STUDIES OF A LOW-NOISE LAMINAR-FLOW TORCH FOR INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY .2. NOISE POWER STUDIES AND INTERFERENCE EFFECTS [J].
DAVIES, J ;
SNOOK, RD .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1987, 2 (01) :27-31
[6]   REDUCTION OF ACOUSTIC NOISE IN ICP EMISSION BY THE ADDITION OF A CHIMNEY [J].
EASLEY, SF ;
MONNIG, CA ;
HIEFTJE, GM .
APPLIED SPECTROSCOPY, 1991, 45 (08) :1368-1371
[7]   NOISE CHARACTERISTICS OF AN INDUCTIVELY COUPLED PLASMA-MASS SPECTROMETER [J].
FURUTA, N ;
MONNIG, CA ;
YANG, PY ;
HIEFTJE, GM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1989, 44 (07) :649-656
[8]   SPATIALLY RESOLVED NOISE AMPLITUDE SPECTRA OF EMISSION SIGNALS FROM AN INDUCTIVELY COUPLED PLASMA [J].
FURUTA, N .
ANALYTICAL SCIENCES, 1990, 6 (05) :683-688
[9]   OPTIMIZATION OF THE MASS SCANNING RATE FOR THE DETERMINATION OF LEAD ISOTOPE RATIOS USING AN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETER [J].
FURUTA, N .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1991, 6 (03) :199-203
[10]   INDUCTIVELY COUPLED PLASMA SOURCE-MASS SPECTROMETRY USING CONTINUUM FLOW ION EXTRACTION [J].
GRAY, AL ;
DATE, AR .
ANALYST, 1983, 108 (1290) :1033-1050