Dose optimization in pelvic radiography by air gap method on CR and DR systems - A phantom study

被引:13
作者
Chan, C. T. P. [1 ]
Fung, K. K. L. [1 ]
机构
[1] Hong Kong Polytech Univ, Dept Hlth Technol & Informat, Kowloon, Hong Kong, Peoples R China
关键词
Air gap method; Dose optimization; Pelvis;
D O I
10.1016/j.radi.2014.11.005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: This study aimed at investigating the feasibility of replacing the anti-scatter grid with an air gap at a pelvic radiographic examination in order to reduce patient dose while retaining diagnostic image quality. Methods: An anthropomorphic pelvis phantom was placed on a device that allowed the adjustment of different air gap thicknesses introduced between the phantom and the image receptor of Computed Radiography (CR) and Digital Radiography (DR) systems. Grid and non-grid images with different air gap thicknesses of both systems were produced. Ovary and testes doses were measured using thermoluminescent dosimeters. Radiographic quality of all images was rated by 5 experienced radiographers blindly using the Image Quality Score (IQS) and Visual Grading Analysis (VGA) systems. Results: Images of diagnostic quality were produced while the grid was replaced by a range of 0-25 cm air gap thickness in the pelvic radiographic examination. At non-grid examination with 10 cm air gap thickness, a maximum of relative dose reduction by 70.7% and 81.6% at CR; 68.6% and 79.4% at DR were achieved respectively at ovary and testes locations of the phantom as compared with their corresponding grid examinations. Conclusion: 10 cm was found to be the optimal air gap thickness at the tested pelvic examination. Effective dose was found to be reduced by 2 and 2.3 times respectively at the CR and DR examinations while the anti-scatter grid was replaced by 10 cm air gap. However, dose reduction effect by air gap method was found to be more pronounced in CR than in DR. (C) 2014 The College of Radiographers. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:214 / 223
页数:10
相关论文
共 45 条
[1]  
Aichinger H., 2011, RAD EXPOSURE IMAGE Q
[2]   DOSE REDUCTION IN RADIOGRAPHY OF THE SPINE IN SCOLIOSIS [J].
ANDERSEN, PE ;
ANDERSEN, PE ;
VANDERKOOY, P .
ACTA RADIOLOGICA-DIAGNOSIS, 1982, 23 (03) :251-253
[3]  
[Anonymous], RADIOGRAPHY
[4]  
BELL N, 2003, RADIOGRAPHY, V0009, P00043
[5]   GRID VERSUS AIR-GAP - A COMPARISON OF CEPHALOMETRIC TECHNIQUES [J].
BENSON, BW ;
FREDERIKSEN, NL ;
GOAZ, PW .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 1994, 77 (01) :86-89
[6]   MONTE-CARLO SIMULATION OF THE SCATTERED RADIATION DISTRIBUTION IN DIAGNOSTIC-RADIOLOGY [J].
BOONE, JM ;
SEIBERT, JA .
MEDICAL PHYSICS, 1988, 15 (05) :713-720
[7]   Increasing film-focus distance (FFD) reduces radiation dose for X-ray examinations [J].
Brennan, PC ;
McDonnell, S ;
O'Leary, D .
RADIATION PROTECTION DOSIMETRY, 2004, 108 (03) :263-268
[8]   Effect of x-ray tube potential on image quality and patient dose for lumbar spine computed radiography examinations [J].
Brindhaban A. ;
Al Khalifah K. ;
Al Wathiqi G. ;
Al Ostath H. .
Australasian Physics & Engineering Sciences in Medicine, 2005, 28 (4) :216-222
[9]  
Bushong SC., 2012, RADIOLOGIC SCI TECHN, V10
[10]  
Carlton R.R., 2012, RADIOGRAPHIC IMAGING